![Eclipse EGA4 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/eclipse/ega4/ega4_user-manual_533846025.webp)
Instruction Manual MM850481 ed. 03a
25
x
Insert the probe in chimney. Stop it using the retaining cone and adjust the immersion depth for a gas sampling in the
middle of the exhaust gas connection pipe.
x
Press the [
ANALYSIS
] (
F1
) key to begin the measurements. The following page will be shown:
O2 20.9 %
CO2 0.0 %
CO 1 ppm
NO 1 ppm
Tf Tc Error
Eff --------
Ref O2
–
GRAPHIC
Pressing the [
W
] and [
X
] keys, it is possible to zoom-in and zoom-out increasing or decreasing the number of displayed
parameters.
O
2
20.9%
CO
2
0.0%
CO 1
ppm
Ref O
2
–
GRAPHIC
O2
20.9%
: CO 2000p
Ta 20.2C
: Eff 0.99
Tf 400.3C
: NO2 100p
K
95.7%
:
CO2 10.2p
:
NOx 500p
:
Ref O
2
–
GRAPHIC
Pressing the [
S
] and [
T
] you can change page for others parameters.
Pressing the [
REF. O
2
] key the measured parameters will be referred to the programmed oxygen value. When selected this
mode, the symbol “
O
2
” will be displayed on the status bar.
O
2
20.9%
CO
2
0.0%
CO 1
ppm
Ref O
2
–
GRAPHIC
If the graphical option (Table E=4 on ordering code) is installed, the instrument can display the measured values as bar-
graph. Press the [
GRAPHIC
] (F2) key to change mode.
O2
CO
K
Ref. O
2
NUMERIC
–
Using GasConfig Software, you can program the graph page: Assign up to 6 different parameter, range and increment
direction. See chapter 9 for software instructions.
Press the [
NUMERIC
] (F2) key to return to the normal display mode.
The integrated CO sensor dilution system, is an automatic device to protect the CO sensor from high concentration of CO.
The device automatically switch to dilution when Automatic dilution procedure is activated (see ch. 9) and the high range
(8000 ppm) value is exceeded.
When dilution is active, the CO value is displayed in reverse and the measuring unit change from ppm to % v/v.