Instruction Manual MM850481 ed. 03a
51
10.3.2 Serial
Port
Selecting this option, you can set the baud rate and the ID number for instrument in infrared serial communication.
SERIAL
Address :
0
Baud-rate
: 9600
ESCAPE
–
Use
the
[
S
] and [
T
] keys to move the cursor symbol to the desired option.
Press the [
ENTER
] key to set a new ID number or baud rate.
Press
[
ESCAPE
] (F2) to return to the previous menu.
10.3.3 Analysis
parameters
Selecting this option you can change the analyzer auto-zeroing time and CH
4
& Natural gas heating probe time (we suggest
to use the pre-set values of 180 sec for autozero time and 60 sec for CH
4
heating time).
ANALYSIS
AutoZ. Time : 180
CH4 Heat. Time: 60
ESCAPE
–
Use
the
[
S
] and [
T
] keys to move the cursor symbol to the desired option.
Press the [
ENTER
] key to set a new Auto-zero time.
Press
[
ESCAPE
] (F2) to return to the previous menu.
10.3.4
P decay time
Selecting this option, you can change the timing for leak test using the decay method. The test has a total period of 310
seconds and the programmable time represent the interval between the measurements.
P decay Time
P1 Time 10
P2 Time 60
P3 Time 120
P4 Time 120
ESCAPE
–
Use
the
[
S
] and [
T
] keys to move the cursor symbol to the desired interval.
Press the [
ENTER
] key to set new acquisition intervals.
Press
[
ESCAPE
] (F2) to return to the previous menu.