TDS-100M Manual 

 

 

6

get lost, even when both the power supply and the backup battery are removed. 

2

The zero point, that is the indicated flow under the situation of motionless liquid, has a great influence 

over the linearity and accuracy of the flow meter. Users should try every possibility to perform the Zero 

Setup that is located at Menu 42. Zero setup should always been performed under the condition of 

motionless liquids. If Zero setup was been performed with motion of the liquid, you can use Menu 43 to 

restore to the original zero point. The zero point will be stored in the RAM memory temporally when you 

use Menu 42. If you want to solidify the zero point, you need to perform Menu 26 “make default paras” 

function again. If this step was omitted, the system will override the RAM zero point by restoring the zero 

point that is stored in the flash memory with next power-on. 

3

The internal calendar is the base of date accumulators. Incorrect date and time will cause messy 

records with the date accumulators. Please the user check the calendar by input correct date and time. The 

calendar runs relying on the backup battery when there is no 24VDC supply. When the battery voltage is 

less than 2 volts, a new backup battery should be replaced with. Replacing the battery with a new one 

should be performed with the 24VDC applied so that the accumulator data and calendar data will not get 

lost. 

4

Please pay great attention to the power supply. The applied voltage should be in the range of +15V to 

24VDC, or 15V to 20VAC. Avoid applying a 110VAC or 220VAC power forever, or the module will be 

damaged. 

5

Pay attention to the display located on Menu 48, which shows the points of lines for the linearity 

correction function. 0 stands for shutting off the function. The default value is 0. Try to make sure that 

value is 0 so as to avoid abnormal flow. 

6

The latest calibrated data which are produced by performing the analog input or output calibration are 

store in the RAM memory temporally. Users need to perform the M26 function to make these data be 

solidified. 

7

Use Menu 49 to check if there are inputs from the RS232 port with network application. If not, please 

check protocol selection software switch located at Menu 63 and the serial port parameters options located 

at Menu62. Select option 1 at Menu 63 for the MOSBUS RTU application. Select option 0 for MODBUS 

ASCII and FUJI extended and Simple Water Meter protocols. 

8

By a flash of the LED indicator for every 2 seconds, TDS-100M means the meter work properly. A 

short flash for every second indicates no-signal or poor signal condition and a long flash for every second 

shows the meter is performing a receiving gain adjusting. 

2. Introduction to TDS-100M 

   

(1) Development introduction   

          The TDS-100M is a compact, high performance general purpose ultrasonic flow meter module. The 
design of the meter is intended to provide for users like system integrators and OEM users with a lowest 
cost but of high performance flow meter. The development of the meter is based on a widely used flow 
meter made by this company. The meter uses only one high performance MSP430 series microprocessor 
made by Taxes Instruments. 

This module can work alone without a LCD and Keypad module. The parameters that the flow meter 

needed to work properly can be (1) setup by a parallel keypad module witch will be plugged onto the 
parallel interface inside of the module, or by a serial keypad module witch will be plugged onto the 

Содержание TDS-100M

Страница 1: ...TDS 100M Manual 1 TDS 100M Ultrasonic Flowmeter User Manual Ver 18 0...

Страница 2: ...DS 100M Manual 2 ground line include upstream ground line and downstream ground line CONTENT 1 Must knows about TDS 100M 5 2 Introduction to TDS 100M 6 3 Menu Window Details 7 4 transducersinstallatio...

Страница 3: ...e software the parameters that TDS 100M needs to work properly could be downloaded by use of the RS232 port on the PC If your computer does not have a RS232 port you need to buy a USB RS232 converter...

Страница 4: ...y located on Menu 48 which shows the points of lines for the linearity correction function 0 stands for shutting off the function The default value is 0 Try to make sure that value is 0 so as to avoid...

Страница 5: ...RS232 and RS485 serial port 4 One or two analog inputs 5 Optional 4 20mA analog output 6 Two OCT output channels 7 Ability to measure sew water 8 4 kinds of different types of transducers can be used...

Страница 6: ...ate positive totalizer adjust the units in M30 M32 03 display instant flow rate negative totalizer adjust the units in M30 M32 04 display instant flow rate date time 05 display heat flow rate total he...

Страница 7: ...is 8 seconds when inputing accures displaying according to the inputing operation when there is no inputing operation it will automatically enter cycle displaying status detailed information in 3 1 4...

Страница 8: ...64 analogue input AI3 65 Analogue input AI4 66 Analogue input AI5 By inputing measuring range the flow meter will turn current signal into data range users need so display related analogue input that...

Страница 9: ...play the Reynolds number and the pipe coefficient diagn osis 95 Display positive negative heat quantity totaliser start cycle display function 0 Display the time of power on off and flow rate 1 Displa...

Страница 10: ...interference function little signal consumable ensure the device work well in longer term so the special signal cable supplied by factory is the best choice for you if using coaxial shielded radio fre...

Страница 11: ...perature and pressure on the installation point should be within the work ability of the transducers 6 Pay more attention to the pipe scaling in inner pipe wall do best to choose the pipe without scal...

Страница 12: ...east above 550mm width is more than D 550 2 mm cement pipe width is more than D 700 2 mm instrument well axial width L is more than D 1200mm when installing transducers avoid the place of flange weldi...

Страница 13: ...ucer installation method Before installation choose density pipe to install transducers and clean the installation area clear away rust paint anti rust layer it is the best to use angle grinder to pol...

Страница 14: ...llation method convenient to use with precise measurement when installation the two transducers horizontally align its centre line is horizontal with pipe axis line suitable pipe diameter range to mea...

Страница 15: ...directly Insertion B type for sement pipe Suitable diameter Above DN80mm Above DN80mm Installation room 550mm 700mm Fluid temperature 40 160 40 160 Transducer material 316L stainless steel 316L stainl...

Страница 16: ...between the vertical side of the fixed position paper and the extended line is A D Starting from A and along the edge of the position paper the length of half of the pipe perimeter is measured and th...

Страница 17: ...ll valves have been welded on the hoops These hoops are directly fixed on the pipe and make sure that the centers of the ball valves overlap A and B respectively Finally the ball valves should be clos...

Страница 18: ...seal the connection place 4 6 8 Length calculation of the part of transducer into the pipe inner wall Insertion style transducer is made of stainless steel by casting As the transducer s length A and...

Страница 19: ...s of the two transducers are face to face that is the holes for wires are in the same direction and either upward or downward 5 After the transducer is installed the screw mut should be screwed tightl...

Страница 20: ...14 20 360 105 75 55 2 14 4 16 25 390 115 85 65 3 14 4 16 32 450 140 100 76 3 18 4 18 40 2 5 500 150 110 84 3 18 4 20 50 200 160 125 100 3 18 4 22 65 200 180 145 120 3 18 4 24 80 225 195 160 135 3 18...

Страница 21: ...l quality is indicated as the Q value display on M90 that represent the receiving signal is good or not TDS 100M series use 00 99 digits to represent signal quality 00 represent the worst signal 99 re...

Страница 22: ...r environment electromagnetic interference or lower receiving signal then the flow rate displayed is poor not be able to work normaly for long time 4 after installation enter M26 to solidify parameter...

Страница 23: ...le to receive signal Loosen contact or not enough couplant between transducer and pipe surface Transducers installed improperly scaling on inner pipe wall is too thick new changed liner Make sure the...

Страница 24: ...K Empty pipe setup in M29 no liquid in pipe or wrong setup if there is liquid actually input 0 value in M29 Attention the codes of Q E displayed do not affect measurement only means current loop and...

Страница 25: ...trument A quotation will be sent to the customer before proceeding with the service 6 3 Software Upgrade Service We provide free of charge software upgrade services Please contact the factory for any...