3, Avoid the installation place of flange,welding line,reducing
4, Enough installation room,convenient for people to stand up to work
§4.4 Quickly input pipe parameter steps:
Need input following parameters when normaly measuring
1. pipe outer diameter
2. pipe wall thickness
3. pipe material
4. liner parameter(if has liner,then include liner thickness and sound velocity)
5. fluid types
6. tansducers tpye(mainframe can support many kinds of
transducers)
7. transducers installation method
8. solidificaiton parameter
§4.5 Clamp on type transducer installation method
:Before installation,choose density pipe to install transducers,and clean the installation
area,clear away rust,paint,anti-rust layer,it is the best to use angle grinder to polish,use cleaning
cloth with alcohol or acetone to clear oil and dust,coat enough couplant around the center of
installation area,attach the transducers on the pipe and fix it without air bubbles or sand between
transducers and pipe wall.
M2 type
Standard S1 type L2 SENSORS
transducer Standard
S1
type
Standard M2
type
Standard L2
type
Suitable pipe
diameter(mm)
DN15-DN100
DN50-DN700
DN300-DN6000
Fluid
temperature
0
℃
~
160
℃
0
℃
~
160
℃
0
℃
~
160
℃
Outer size
45×30×30mm
60×45×45
80×70×55
quality 75g 250g 650g
Remarks:users input transducers parameters by yourself