background image

TDS-100M Manual 

 

 12

-9 

PT100 setup zero point at higher temperature

>55

 

-A 

PT100 standard calibration at 50

 

-B 

PT100 standard calibration at 84.5

 

 

Note

 

means common used menus,

red color

 means new added or changed 

functions,

blue color 

means the menus related with heat quantity measurement 

 

 

4.transducers installation 

§4.1 Unpack checking 

Check whether the spare parts comply with packing list,enclosure is broken or not during 
transportation?did screw drop?connecting line is loose or not?if have questions,pls cotact factory. 

§4.2 Power supply and cable 

When users place an order,pls inform factory what kinds of power supply is 

needed,normaly the power supply of the flow meter have following options: 

: DC24V or DC8~36V 

Transducers signal cable of TDS-100M series ultrasonic flow meter adopts high frequency special 

shielded twisted pair cable.because sending and receiving circuit adopts balanced transmitting 

and balanced receiving principle.so to increase the anti-interference function,little signal 

consumable ,ensure the device work well in longer term.so the special signal cable supplied by 

factory is the best choice for you.if using coaxial shielded radio frequency cable or poor quality 

twisted pair cable,it could lower accuracy and the function of the device.when there is bigger 

interference signal from outside,the device could not measure normally. 

§4.3 Required installation condition 

Installation of The new TDS-100M series is the easiest and convenient way in the 

installation of all flow meters,just choosing a suitable measurement point,input the pipe 
parameters of this pipe point to the flow meter,then fix the transducers on the pipe. 

§4.3.1 choosing measurement point 

To ensure measurement accuracy and stability,the installation point of tranducers should  be on 
the straight pipe full of well distributed fluid(when installing,the pipe must be full of liquid),comform 
to following principle: 
1.Pipe must be full of liquid that is uniform and easy to travel the ultrasonic beam(vertical pipe or 
horizontal pipe) 

 

2.Upstream transducer should be installed at the place where the upstream length of the straight 

Содержание TDS-100M

Страница 1: ...TDS 100M Manual 1 TDS 100M Ultrasonic Flowmeter User Manual Ver 18 0...

Страница 2: ...DS 100M Manual 2 ground line include upstream ground line and downstream ground line CONTENT 1 Must knows about TDS 100M 5 2 Introduction to TDS 100M 6 3 Menu Window Details 7 4 transducersinstallatio...

Страница 3: ...e software the parameters that TDS 100M needs to work properly could be downloaded by use of the RS232 port on the PC If your computer does not have a RS232 port you need to buy a USB RS232 converter...

Страница 4: ...y located on Menu 48 which shows the points of lines for the linearity correction function 0 stands for shutting off the function The default value is 0 Try to make sure that value is 0 so as to avoid...

Страница 5: ...RS232 and RS485 serial port 4 One or two analog inputs 5 Optional 4 20mA analog output 6 Two OCT output channels 7 Ability to measure sew water 8 4 kinds of different types of transducers can be used...

Страница 6: ...ate positive totalizer adjust the units in M30 M32 03 display instant flow rate negative totalizer adjust the units in M30 M32 04 display instant flow rate date time 05 display heat flow rate total he...

Страница 7: ...is 8 seconds when inputing accures displaying according to the inputing operation when there is no inputing operation it will automatically enter cycle displaying status detailed information in 3 1 4...

Страница 8: ...64 analogue input AI3 65 Analogue input AI4 66 Analogue input AI5 By inputing measuring range the flow meter will turn current signal into data range users need so display related analogue input that...

Страница 9: ...play the Reynolds number and the pipe coefficient diagn osis 95 Display positive negative heat quantity totaliser start cycle display function 0 Display the time of power on off and flow rate 1 Displa...

Страница 10: ...interference function little signal consumable ensure the device work well in longer term so the special signal cable supplied by factory is the best choice for you if using coaxial shielded radio fre...

Страница 11: ...perature and pressure on the installation point should be within the work ability of the transducers 6 Pay more attention to the pipe scaling in inner pipe wall do best to choose the pipe without scal...

Страница 12: ...east above 550mm width is more than D 550 2 mm cement pipe width is more than D 700 2 mm instrument well axial width L is more than D 1200mm when installing transducers avoid the place of flange weldi...

Страница 13: ...ucer installation method Before installation choose density pipe to install transducers and clean the installation area clear away rust paint anti rust layer it is the best to use angle grinder to pol...

Страница 14: ...llation method convenient to use with precise measurement when installation the two transducers horizontally align its centre line is horizontal with pipe axis line suitable pipe diameter range to mea...

Страница 15: ...directly Insertion B type for sement pipe Suitable diameter Above DN80mm Above DN80mm Installation room 550mm 700mm Fluid temperature 40 160 40 160 Transducer material 316L stainless steel 316L stainl...

Страница 16: ...between the vertical side of the fixed position paper and the extended line is A D Starting from A and along the edge of the position paper the length of half of the pipe perimeter is measured and th...

Страница 17: ...ll valves have been welded on the hoops These hoops are directly fixed on the pipe and make sure that the centers of the ball valves overlap A and B respectively Finally the ball valves should be clos...

Страница 18: ...seal the connection place 4 6 8 Length calculation of the part of transducer into the pipe inner wall Insertion style transducer is made of stainless steel by casting As the transducer s length A and...

Страница 19: ...s of the two transducers are face to face that is the holes for wires are in the same direction and either upward or downward 5 After the transducer is installed the screw mut should be screwed tightl...

Страница 20: ...14 20 360 105 75 55 2 14 4 16 25 390 115 85 65 3 14 4 16 32 450 140 100 76 3 18 4 18 40 2 5 500 150 110 84 3 18 4 20 50 200 160 125 100 3 18 4 22 65 200 180 145 120 3 18 4 24 80 225 195 160 135 3 18...

Страница 21: ...l quality is indicated as the Q value display on M90 that represent the receiving signal is good or not TDS 100M series use 00 99 digits to represent signal quality 00 represent the worst signal 99 re...

Страница 22: ...r environment electromagnetic interference or lower receiving signal then the flow rate displayed is poor not be able to work normaly for long time 4 after installation enter M26 to solidify parameter...

Страница 23: ...le to receive signal Loosen contact or not enough couplant between transducer and pipe surface Transducers installed improperly scaling on inner pipe wall is too thick new changed liner Make sure the...

Страница 24: ...K Empty pipe setup in M29 no liquid in pipe or wrong setup if there is liquid actually input 0 value in M29 Attention the codes of Q E displayed do not affect measurement only means current loop and...

Страница 25: ...trument A quotation will be sent to the customer before proceeding with the service 6 3 Software Upgrade Service We provide free of charge software upgrade services Please contact the factory for any...

Отзывы: