HEALTH IMAGING
© Eastman Kodak Company
Publication No. 8B7489
01MAY97
Supersedes 8B7489
September 1996
MODIFICATION INSTRUCTIONS
for the
Kodak X-Omat MULTILOADER 300
Service Code: 3058
Modification No. 35
Kodak X-Omat 270 RA PROCESSOR
Service Code: 3059
Modification No. 21
Kodak X-Omat 180 LP and 180 LPS PROCESSORS
Service Codes: 3227, 3198
Modification No. 8
Type 1 Selective
Purpose:
To improve the reliability of the PROCESSOR by:
• moving the connections for the DRYER OVER-TEMPERATURE and the DRYER
THERMISTOR to outside of the TANK, to reduce corrosion of the
CONNECTORS
• installing a new YOKE AND PIN ASSEMBLY, to minimize the possibility of diluting the fixer
• installing plastic SPROCKETS in the DRYER RACK, to eliminate corrosion of the SPROCKETS
Содержание 3059
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 12: ......
Страница 13: ......
Страница 14: ......
Страница 15: ......
Страница 16: ......
Страница 17: ......
Страница 18: ......
Страница 19: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 27: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 69: ...MODIFICATION INSTRUCTIONS TYPE1 10 29JUN96 8B6658 ...
Страница 107: ...mb3058_43_15mar 99 fm Health Imaging EASTMAN KODAK COMPANY ROCHESTER N Y 14650 Kodak and X Omat are trademarks ...