12.2 Measuring
1. Before you start a measurement, check that
- you have selected the proper measuring mode (see section 9.1).
- you have assigned the proper method (see section 11.1.2).
- you have assigned the proper sample ID, if needed (see
- the sample’s temperature is between 0 °C and 100 °C (32 °F and
212 °F).
- suitable solvents for cleaning are available.
2. Fill the sample (see section 12.1.1 for filling with the filling tube or
section 12.1.2 for filling with a syringe).
The instrument shows a continuous reading of the current measuring
values.
3. Check for bubble-free filling and that the measuring cell is fully filled.
Possible reasons for bubbles in the measuring cell are:
- gas bubbles in the sample
- leaky connection of the filling tube, the pump or the screw plug/
syringe adapter
4. Tap the
key to start a measurement.
The result is displayed and stored in the internal data memory.
5.
Filled with filling tube
: Place the filling tube in a suitable waste vessel
and press the pump lever to empty the measuring cell.
Filled with syringe
: Unplug the syringe to let the sample drain off via the
filling tube.