34
EN
10 Checks and Adjustments
Perform a water check every day before you start the measurements to verify
that the instrument is measuring with adequate accuracy.
Perform a water adjustment if the water check advises you to do so.
10.1 Performing a Water Check
Use freshly degassed ultra-pure (bi-distilled or deionized) water.
Performing a water check
1. Clean the measuring cell as described in section 14.
2. Rinse the measuring cell until no more cleaning agent residues are
present.
3. In the quick access area, activate
("Check" function).
4. Select "Water Check" and tap <OK>.
5. Follow the instructions on the screen.
6. After the procedure has finished, the check results are displayed.
If the deviations from the target values exceed defined limits, you will see
a recommendation to perform a water adjustment.
Tap <OK>.
7. Tap <Back> to return to the main screen.
10.2 Performing a Water Adjustment
Use freshly degassed ultra-pure (bi-distilled or deionized) water.
Performing a water adjustment
1. Clean the measuring cell as described in section 14.
2. Rinse the measuring cell until no more cleaning agent residues are
present.
3. Tap <Menu> and select
Adjustments > Water Adjustment.
4. Fill ultra-pure water (15 °C to 25 °C) into the instrument.
5. Tap <Start>.