2.2 Installation and Use
• The installation procedure shall be carried out only by authorized
personnel who are familiar with the installation instructions.
• Use only accessories and consumables supplied or approved by
Eagle
Eye
• Ensure that all operators have been trained beforehand to use the
instrument safely and correctly.
• Ensure that the instrument is sufficiently supervised during operation.
• In case of damage or malfunction, do not continue operating the
instrument. Do not operate the instrument under conditions which could
result in damage to goods or injuries or loss of life.
• Do not expose the instrument to temperatures below 0
°C (32
°F) when
the measuring cell contains water. Freezing water will cause rupture of
the measuring cell.
• The instrument is not insulated against high voltages. Measuring
samples under high voltage (e.g. in energized battery banks) bears the
risk of an electric shock. Define appropriate testing procedures and
safety measures to protect yourself from any electric shock.
Operation in areas with risk of explosion
• The instrument is
not
explosion-proof and therefore must not be
operated in areas with risk of explosion.
General precautions
• Observe and adhere to your national safety regulations regarding the
handling of all substances associated with your measurements (e.g. use
safety goggles, gloves, respiratory protection, etc.).
• Before a measurement, check the wetted parts of the instrument for
chemical resistance to the samples and cleaning agents used.