31
EN
9
Measuring Settings
9.1
Setting the Measurement Mode
SG-Ultra Max
features three measurement modes, "Precise", "Fast" and
"Manual".
For each measurement mode, different stability criteria have to be fulfilled
before the measuring result is stored. The stability criterion is always related
to the temperature:
- Measurement mode "Precise":
The result is stored as soon as the measured temperature value
stays within 0.2 K for 10 seconds. This measurement mode delivers
the most accurate results, but may take a longer time in case the
sample temperature differs greatly from the ambient temperature.
- Measurement mode "Fast":
The result is stored as soon as the measured temperature value
stays within 0.4 K for 10 seconds. This measurement mode delivers
quicker results than the "Precise" mode, but as the density is highly
temperature-dependent, the measured result is not as accurate.
- Measurement mode "Manual":
You decide yourself when to store your measurement result. The
result is stored immediately after tapping the
key.
Setting the measurement mode
1. Tap <Menu> and select
Setup > Measurement Mode
.
2. Tap <Edit>.
3. Select one of three measurement modes:
- Precise
- Fast
- Manual
4. Tap <Back> repeatedly to return to the main screen.
Содержание SG-Ultra Max
Страница 1: ...Instruction Manual SG Ultra Max Portable Density Meter...
Страница 6: ...6 EN...