12.1.2 Filling with a Syringe
IMPORTANT:
When delivered, the instrument is equipped with the pump
lever without lock function. Replace this pump lever by the pump lever with
lock function (supplied with the instrument) as described in appendix 15.
1. Mount the syringe adapter (see section 6.2).
2. Lock the pump lever.
Push the pump lever down and forward simultaneously to minimize the
dead volume in the syringe plunger.
3. Place the filling tube in a suitable waste vessel.
4. Fill the syringe with the sample.
5. Attach the syringe to the syringe adapter and fill the measuring cell (see
Fig. 12-2: Filling sample using a syringe
WARNING
If you remove the pump when filling with a syringe, or remove the syringe
after filling, the system becomes leaky. Harmful liquids leaking from the
instrument may cause injuries.
• Do not remove the pump.
• Do not remove the syringe during measurement.