Eagle Eye SG-1000C User Manual 

Trouble? 

If abnormal measurements are experienced, 
It may be due to low battery power. Replace 
with new alkaline battery. 

CAUTION 

Care should be taken to avoid damage to the prism. Use soft tissue paper when
wiping the prism surface to avoid scratching. Never make contact with any metal
tools or abrasive materials.

The instrument is designed to measure the electrolyte solution of lead acid
batteries. Do not use the measurement of other electrolyte solutions such as in
alkaline batteries.

The electrolyte contains sulfuric acid and is toxic. Burns may occur if contact is
made on the skin or clothes. Blindness may occur if contact is made to the eyes.
Always wear protective glasses and rubber gloves. If the electrolyte is splashed
onto the skin or clothing, rinse with sufficient amounts of cold water. If it gets
into an eye, flush with cold clean water for about 15 minutes and seek immediate
medical attention.

Remove the battery from the Unit for long periods of non-use and storage.

Temperature 

Attention to the following points is needed during the measurement when the 
temperature of the electrolyte varies greatly with the ambient temperature.  

The instrument compensates for the temperature of the sample automatically 
according to the temperature detected on the prism in the Sample Chamber. 

The 

instrument always compensates to a temperature of 68°F (20°C). 

If the temperature of 

the sample is different from the prism temperature, wait approximately 3 seconds for 
every 2 degree F (3 seconds for every 1 degree C) of difference between temperatures 
before pressing the Start Switch. (If any difference between temperatures is suspected, 
wait some time before pressing the Start Switch.) 

Содержание SG-1000

Страница 1: ...SG 1000 Specific Gravity Tester User Manual The BEST Battery Testing Equipment The BEST Customer Support PERIOD www eepowersolutions com Copyright 2012 Eagle Eye Power Solutions LLC V1 0...

Страница 2: ......

Страница 3: ...shed onto the skin or clothing rinse with sufficient amounts of cold water If it gets into an eye flush with cold clean water for about 15 minutes and seek immediate medical attention Remove the batte...

Страница 4: ...mes and Functions of Parts 4 2 Inserting the Nozzle 6 3 Inserting the Battery 7 4 Attaching the Wrist Strap 8 5 Zero Setting 9 6 Measurement of Sample 11 7 Explanation of Error Messages 13 8 Temperatu...

Страница 5: ...Eagle Eye SG 1000C User Manual 4 1 Names and Functions of Parts...

Страница 6: ...setting 4 START OFF key Press to take measurement To turn off the instrument press and hold down the key for 2 seconds 5 Battery Cover Remove to replace battery 6 Nozzle Cap Used to protect tip of noz...

Страница 7: ...e is packed separately from the Main unit Insert the Nozzle to Main Unit following the procedure described below 1 Push the Nozzle into the hole located 2 Then fit the nozzle into the slit by at the b...

Страница 8: ...itself is not guaranteed and be sure to use a new battery if its voltage is low 3 Close the Battery Cover and using a Screwdriver turn the locking pin to the CLOSE position Note The SG 1000 does not h...

Страница 9: ...SG 1000C User Manual 8 4 Attaching the Wrist Strap Attach the wrist strap to the hook located on the lower side of the Main Unit When measuring samples loop the strap around your wrist to avoid any a...

Страница 10: ...at the Sample Chamber has been filled with water If bubbles are present in the Sample Chamber operate the Tube Pump again to refill water into the Sample Chamber 2 Dip the tip of the Nozzle in water a...

Страница 11: ...Chamber Allow the unit to equilibrate to the ambient temperature before taking measurements Zero setting can be performed anytime Using water the unit will be calibrated to zero Use distilled water fo...

Страница 12: ...d squeeze and release the Tube Pump a few times to draw the sample into the sample chamber Once the sample is drawn into the unit Drain it completely and then refill the sample into the unit again Thi...

Страница 13: ...er for a few days the prism surface will be soiled and consequently cause errors during measurement To clean the prism surface use a soft tissue soaked with ethyl or isopropyl alcohol If the electroly...

Страница 14: ...when performing zero set LLL Sampling error Electrolyte in the Sample Chamber is insufficient to cover the prism surface or air bubbles may be present HHH Range over error The sample measured is out o...

Страница 15: ...ce 1 To replace the Tube Pump slide it to the left to release and replace it with a new one 2 After measurement fill and drain the Sample Chamber with distilled or tap water several times to clean Com...

Страница 16: ...The Sample Chamber Sample Hook and or Tube Pump may have been freed from the normal position Make sure the Sample Chamber Hook is properly attached to the unit Adjust the Sample Chamber and Tube Pump...

Страница 17: ...Eagle Eye SG 1000C User Manual 16...

Страница 18: ...the ambient temperature and the sample temperature close to equilibrium Does the battery need replacement Is the O ring damaged in the Sample Chamber Is the Tube Pump damaged Is the Nozzle damaged Is...

Страница 19: ...ttery Refer to page 8 on how to replace the battery and other remarks The unit uses a 9V DC battery 006P Is the O ring damaged in the Sample Chamber Remove the Sample Chamber cover from the unit see p...

Страница 20: ...the Nozzle is securely attached Leakage of liquid or air can be observed if the Tube Pump Nozzle or cover has been damaged or the Nozzle is loose If damage is found replace the damaged part Two spare...

Страница 21: ...he sample before pressing the START key 11 Specifications Model SG 1000 Detection method Optical refraction Measurement range Specific gravity of electrolyte 1 000 to 1 300 Resolution 0 001 Measuremen...