background image

  

 

Eagle Eye SG-1000C User Manual 

 

 

 

12 

 

(4) 

The displayed value will remain on 

the LCD for 5 minutes.  
 

(5) 

To reset the display, press and hold 

down the START switch for 2 seconds.  
 

(6) 

Before storing the SG1000F, squeeze 

and release the Tube Pump several 
times to drain the sample from the 
Chamber. Then, repeat filling and 
draining the Sample Chamber several 
times with water. Drain the water well 
before storing.  
 
 
 
 
 
 
 
 
 
 
 
 

 

 

If electrolyte is left in the Sample Chamber for a few days, the prism surface will be 
soiled and consequently cause errors during measurement. To clean the prism 
surface, use a soft tissue soaked with ethyl or isopropyl alcohol. If the electrolyte is 
left in the Sample Chamber for a long period of time, corrosion of the prism occurs 
and will cause permanent damage to the unit.  

 

The Sample Chamber should be shaded when taking measurements. Erroneous 
measurements may result if the Sample Chamber is exposed to too much external 
light.  

 

When sample liquid cannot be removed from the Nozzle by operating the Tube 
Pump, apply tissue paper to the tip of Nozzle and manually squeeze it out.  

 

Содержание SG-1000

Страница 1: ...SG 1000 Specific Gravity Tester User Manual The BEST Battery Testing Equipment The BEST Customer Support PERIOD www eepowersolutions com Copyright 2012 Eagle Eye Power Solutions LLC V1 0...

Страница 2: ......

Страница 3: ...shed onto the skin or clothing rinse with sufficient amounts of cold water If it gets into an eye flush with cold clean water for about 15 minutes and seek immediate medical attention Remove the batte...

Страница 4: ...mes and Functions of Parts 4 2 Inserting the Nozzle 6 3 Inserting the Battery 7 4 Attaching the Wrist Strap 8 5 Zero Setting 9 6 Measurement of Sample 11 7 Explanation of Error Messages 13 8 Temperatu...

Страница 5: ...Eagle Eye SG 1000C User Manual 4 1 Names and Functions of Parts...

Страница 6: ...setting 4 START OFF key Press to take measurement To turn off the instrument press and hold down the key for 2 seconds 5 Battery Cover Remove to replace battery 6 Nozzle Cap Used to protect tip of noz...

Страница 7: ...e is packed separately from the Main unit Insert the Nozzle to Main Unit following the procedure described below 1 Push the Nozzle into the hole located 2 Then fit the nozzle into the slit by at the b...

Страница 8: ...itself is not guaranteed and be sure to use a new battery if its voltage is low 3 Close the Battery Cover and using a Screwdriver turn the locking pin to the CLOSE position Note The SG 1000 does not h...

Страница 9: ...SG 1000C User Manual 8 4 Attaching the Wrist Strap Attach the wrist strap to the hook located on the lower side of the Main Unit When measuring samples loop the strap around your wrist to avoid any a...

Страница 10: ...at the Sample Chamber has been filled with water If bubbles are present in the Sample Chamber operate the Tube Pump again to refill water into the Sample Chamber 2 Dip the tip of the Nozzle in water a...

Страница 11: ...Chamber Allow the unit to equilibrate to the ambient temperature before taking measurements Zero setting can be performed anytime Using water the unit will be calibrated to zero Use distilled water fo...

Страница 12: ...d squeeze and release the Tube Pump a few times to draw the sample into the sample chamber Once the sample is drawn into the unit Drain it completely and then refill the sample into the unit again Thi...

Страница 13: ...er for a few days the prism surface will be soiled and consequently cause errors during measurement To clean the prism surface use a soft tissue soaked with ethyl or isopropyl alcohol If the electroly...

Страница 14: ...when performing zero set LLL Sampling error Electrolyte in the Sample Chamber is insufficient to cover the prism surface or air bubbles may be present HHH Range over error The sample measured is out o...

Страница 15: ...ce 1 To replace the Tube Pump slide it to the left to release and replace it with a new one 2 After measurement fill and drain the Sample Chamber with distilled or tap water several times to clean Com...

Страница 16: ...The Sample Chamber Sample Hook and or Tube Pump may have been freed from the normal position Make sure the Sample Chamber Hook is properly attached to the unit Adjust the Sample Chamber and Tube Pump...

Страница 17: ...Eagle Eye SG 1000C User Manual 16...

Страница 18: ...the ambient temperature and the sample temperature close to equilibrium Does the battery need replacement Is the O ring damaged in the Sample Chamber Is the Tube Pump damaged Is the Nozzle damaged Is...

Страница 19: ...ttery Refer to page 8 on how to replace the battery and other remarks The unit uses a 9V DC battery 006P Is the O ring damaged in the Sample Chamber Remove the Sample Chamber cover from the unit see p...

Страница 20: ...the Nozzle is securely attached Leakage of liquid or air can be observed if the Tube Pump Nozzle or cover has been damaged or the Nozzle is loose If damage is found replace the damaged part Two spare...

Страница 21: ...he sample before pressing the START key 11 Specifications Model SG 1000 Detection method Optical refraction Measurement range Specific gravity of electrolyte 1 000 to 1 300 Resolution 0 001 Measuremen...

Отзывы: