Symptom
Remedy
The unit will not turn on. The contact plate
in the battery compartment may have
been bent creating a gap between the
plate and the battery.
Restore contact between the contact plate
and battery by using a screwdriver to
return the plate to the normal position.
The Sample Chamber, Sample Hook
and/or Tube Pump may have been freed
from the normal position.
Make sure the Sample Chamber Hook is
properly attached to the unit. Adjust the
Sample Chamber and Tube Pump into
position and lock in place with the Hook.
Unit is cracked.
Any cracks in the casing cannot be
repaired. Return the unit to the
manufacturer.
The SG-1000 is a precision optical instrument and should be handled with care. Do not
drop or subject to any heavy impact or shock.