Eagle Eye SG-1000C User Manual
18
*Is the power level of the
battery low?
The unit does not measure if the battery
is low. The battery level can be checked
by using a multimeter. If the mattery is
low, replace it with a new battery. (Refer
to page 8 on how to replace the battery
and other remarks.)
The unit uses a 9V DC battery (006P)
*Is the O-ring damaged in the
Sample Chamber?
Remove the Sample Chamber cover
from the unit (see page 4), and check
the O-ring for deformation or damage.
If the O-ring has been damaged replace
it with a new O-ring which can be
purchased from your dealer.
After replacing the O-ring, properly
replace the Sample Chamber cover.