Eagle Eye SG-1000C User Manual
10
(4)
Press the ZERO key. “000” will blink
three times and “000” will be displayed.
(5)
Zero-Setting is now complete.
Remove the Nozzle from the water and
squeeze the Tube Pump to drain the
water from the Sample Chamber.
Allow the unit to equilibrate to the ambient temperature before taking
measurements.
Zero-setting can be performed anytime. Using water, the unit will be calibrated to
zero.
Use distilled water for zero-setting. Tap water may be used if the impurity level of
the water is less than 0.01%.
Before zero-setting, the water used should also be at the ambient temperature.
Zero-setting should be performed whenever the battery is replaced.