flash back and forth between “FAn” and “5”. With the display function turned ”Off”, the 
display does not show the gas concentration, but will show “FAn” as long as the fan 
relay is activated 

Note: If the Fan relay does not close within 2 minutes, consider  these 
possibilities: 

a. Gas cylinder is empty, check the pressure gauge. Replace the gas cylinder if 25psi or
less.
b. Unit needs to be re-calibrated (go through recalibration and re-test).
c. Detector is in need of servicing (return unit to factory for servicing).
d. Detector has fan relay set to disable (OFF) or 20% LEL. Set fan relay to 5% LEL and
repeat the test.
7. Remove the gas from the sensor. Proceed to test the alarm relay or replace the top
cover.

Testing the Alarm Relay 
Note: The gas concentration to activate the Alarm relay depends on the setting. 

Connect the 20% LEL cylinder of Combustible Gas to the regulator. Ensure that the gas 
used for calibration matches the gas selected  in the Eagle Eye GD-3000  Combustible 
Gas Detector configuration. 

1. Check the pressure gauge. If there is 25psi or less the cylinder should be replaced.
2. Place the test hood over the gas sensor. Turn on the regulator to start the gas flow.
3. The Fan relay should activate according to the settings.
4. With the display function turned “On” and the gas concentration reaching the Alarm
Relay setting, (20% LEL, for example) the display will flash back and forth between
“ALr” and “20”. The buzzer will sound indicating “Alarm” if the buzzer is turned “On”.
With the display function turned off the display does not show the gas concentration, but
will show “ALr” when the Alarm relay is activated.

5.

Note: If the Alarm relay fails to operate within 2 minutes, consider these

possibilities:

a. Gas cylinder is empty, check the pressure gauge. Replace the gas cylinder if 25psi or
less.
b. Unit needs to be re-calibrated (go through recalibration and re-test).
c. Detector is in need of servicing (return unit to factory for servicing).
d. Detector has Alarm relay set to disable (OFF). Set Alarm relay to 20% LEL and
repeat the test.
6. Remove the gas from the sensor after test. Proceed to test the 4-20mA output or
replace the top cover.

20 

Содержание GD-3000

Страница 1: ...Eagle Eye GD 3000 User Manual Updated 10 09 13 4031 W Kiehnau Ave Milwaukee WI 53209 TEL 1 414 962 3377 TOLL FREE 1 877 805 3377 FAX 1 414 962 3660 www eepowersolutions com V1 0...

Страница 2: ...ormation 7 3 2 Location 7 3 3 Installation 7 3 4 Power Up 11 3 5 Operation 12 3 6 Default Configuration Factory Settings 12 3 7 Onboard Diagnostics 15 3 8 Sensor Poisons 16 4 MAINTENANCE 16 4 1 Cleani...

Страница 3: ...untrained or unqualified persons or use that is not in accordance with these User Instructions may adversely affect product performance and result in sickness or death Use only for monitoring the gas...

Страница 4: ...ibration gas at the required concentration level Do not calibrate with expired calibration gas If the instrument cannot be calibrated do not use until the reason can be determined and corrected Do not...

Страница 5: ...GD 3000 Combustible Gas Detector uses a microcomputer controlled electronic system to measure the concentration of combustible gas actuate relays and provide a 4 20mA 0 50 LEL output The GD 3000 Combu...

Страница 6: ...ectable at 1 2 3 4 5 6 7 8 9 10 default 11 12 13 14 15 16 17 18 19 20 LEL Fan Delay Settings of 0 1 3 default 5 and 10 minutes Fan Minimum Run Time settings are OFF default 3 5 10 or 15 minutes Alarm...

Страница 7: ...the GD 3000 Combustible Gas Detector where the normal ambient temperature is below below 0 F or exceeds 125 F 18 C or above 52 C If the target gas is lighter than air natural gas Methane or Hydrogen...

Страница 8: ...that the disable setting will cause the alarm relay not to engage at all 7 The dry contact SPDT fan relay has three terminals The common COM normally open N O and the normally closed N C contact The f...

Страница 9: ...Gas Detector high on a wall about one foot down from the ceiling If the gas used is propane LP mount the GD 3000 Combustible Gas Detector on a wall or column one foot above the floor 3 See applicable...

Страница 10: ...cked up power to Eagle Eye GD 3000 Combustible Gas Detector 3 Eagle Eye GD 3000 Combustible Gas Detector dry contact fan relay can use panel power or fan power Check ratings on panel fan 4 Match fan r...

Страница 11: ...ute the test cycle any time power is dropped and reapplied i e power failure During the self test cycle the unit will display the firmware version number then count down from 60 to 0 and finally go in...

Страница 12: ...FAn and ALr and sound as outlined above 3 6 Default Configuration Factory Settings The default Display setting is Off The default Power Up Test setting is On The default Buzzer setting is Off The def...

Страница 13: ...is displayed then press Enter to return to normal operation Selecting Display Option dSP To select the Display Configuration in normal mode press the Next button twice to get to Con or the Configurati...

Страница 14: ...e FrS or Fan Relay setting Press Next six times to get to FrS then press Enter If the fan relay is OFF disabled press Next to change it to 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 or 20 LEL fla...

Страница 15: ...d verify unit operations If a problem is found the unit will switch to a fail safe error mode or trouble condition In this error mode the Fan and Alarm relays will be activated the 4 20mA current loop...

Страница 16: ...er serviceable parts and substitution of components may impair intrinsic safety which may adversely affect product performance and result in sickness or death CAUTION Avoid the use of harsh cleaning m...

Страница 17: ...er to the General section of this User Instruction 6 Locate the switch labeled ENTER TEST on the left side of the printed circuit board Press the Test switch once 7 The Eagle Eye GD 3000 Combustible G...

Страница 18: ...or calibration verification test bump test to ensure proper performance of the monitor Failure to do so may adversely affect product performance and result in sickness or death When performing a cali...

Страница 19: ...the regulator Ensure that the gas used for calibration matches thegas selected in the Eagle Eye GD 3000 Combustible Gas Detector configuration 3 Assemble regulator hose and test hood and place the tes...

Страница 20: ...00 Combustible Gas Detector configuration 1 Check the pressure gauge If there is 25psi or less the cylinder should be replaced 2 Place the test hood over the gas sensor Turn on the regulator to start...

Страница 21: ...vicing return unit to factory for servicing d Detector has 4 20mA option set to OFF Set 4 20mA option to On and repeat the test 6 Remove the gas from the sensor Re assemble the Eagle Eye GD 3000 Combu...

Страница 22: ...may be present and could result in overexposure and cause sickness or death 4 3 1 Zero the Sensor 1 Remove the Philips screw on the front of the Eagle Eye GD 3000 Combustible Gas Detector Pull the fro...

Страница 23: ...and the calibration will progress The display will show this for a maximum of 165 seconds 10 When the calibration is successful the display will flash back and forth between 10 and PA S 11 Remove the...

Страница 24: ...MPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE ARE LIMITED TO A PERIOD OF TWO 2 YEARS FROM THE PURCHASE DATE Eagle Eye shall not be liable for any incidental or consequential...

Отзывы: