53
TECHNICAL CHARACTERISTICS
y
GUIDANCE AND MANUFACTURER’S DECLARATION
RADIO FREQUENCY EMISSIONS
is equipped with a radio-frency transmitter at 13.56 MHz in an ASK modulation, and at a power lower than 42 dB(µA/m) à 10m.
ELECTROMAGNETIC EMISSIONS
is intended for use in the electromagnetic environment specified below. The user of
should assure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
uses RF energy only for its internal function. Therefore, its radiated
emissions are very low and are not likely to cause interference in nearby electronic
equipement.
RF emissions - CISPR 11
Class B
is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low voltage power supply network that
supplies buildings.
Harmonic emissions - IEC 61000-3-2
Class A
Voltage fluctuations
Flicker emissions - IEC 61000-3-3
Passed
ELECTROMAGNETIC IMMUNITY
is intendes for use in the electromagnetic environment specified below. The user of
should assure that it is used in such
an environment.
Immunity test
IEC60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6KV contact
± 8kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient
/ burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 2KV for power supply lines.
Not applicable.
Mains power quality should be that of a typical public low
voltage power supply network.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical public low
voltage power supply network.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines.
IEC 61000-4-11
<
5 % UT (
>
95% dip in UT)
for 0,5 cycle.
40 % UT (60% dip in UT)
for 5 cycles.
70 % UT (30% dip in UT)
for 25 cycles.
<
5 % UT (
>
95% dip in UT)
for 5 seconds.
<
5 % UT (
>
95% dip in UT)
for 0,5 cycle.
40 % UT (60% dip in UT)
for 5 cycles.
70 % UT (30% dip in UT)
for 25 cycles.
<
5 % UT (
>
95% dip in UT)
for 5 seconds.
Mains power quality should be that of a typical public low
voltage power supply network. If the user of
requires continued operation during power mains
interruption, it is recommended that
be powered
from an uninterruptible power supply or a battery.
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristics of a typical location connected to public
low voltage power supply network.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
de 150Hz to 80MHz
3V/m
de 80 MHz to 2,5 GHz
3Vrms
3V/m
Portable and mobile RF communications equipment
should be used no closer to any part of
, including
cables, than the recommended separation of distance
calculated from the equation applicable to the frequency
of transmitter:
d = 1,2 √P.
d = 1,2 √P 80MHz to 800 MHz.
d = 2,3 √P 800 MHz to 2,5 GHz.
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strength from fixed RF tranmitters, as
determined by an electromagnetic field survey
a
should
be less than the compliance level in each frequency
range
b
.
Interference may occur in the vicinity of equipment
matked with the following symbol: