18
Operating instructions for EE360 Moisture in Oil Transmitter
HttP-Webserver (Port 80)
For a quick communication check you can enter the desired IP in a web browser to get a brief
system status.
Alternatively you can send an ICMP echo request (“ping”) to check correct communication set-
tings.
EE36-over-UDP (Port 5234)
Backwards compatible with EE36 series.
Communication via Ethernet takes place by means of UDP packets, which transport a command
or a command response in EE36 protocol format as payload data.
See “Ethernet - communication protocol” which can be found on https://www.epluse.com/en/ser-
vice-support/download-center/
4.4.2 IPv4-Settings
Factory setting:
IP-Address: 192.168.0.64
Subnet-Mask: 255.255.255.0
Gateway: 192.168.0.1
DNS: 192.168.0.1
DHCP options can be set with jumper on the PCB
Factory setting: DHCP disabled (static IP)
Fig. 18
Ethernet Module - DHCP setting
The.setup.of.the.Modbus.TCP.communication.can.be.performed.via.EE-PCS.Product.Config-
uration Software or via display and push buttons.
Modifying the IP-Address via EE-PCS or display is possible only when the DHCP jumper is con-
figured.“Static”..Otherwise.the.IP-settings.are.read-only