Operating instructions for EE360 Moisture in Oil Transmitter

1  General

This manual is included in the scope of supply and serves to ensure proper handling and  

optimum performance of the instrument. The manual shall be read before putting the device into opera-

tion. The manual is relevant for all staff involved in transport, setup, operation, maintenance and repair. 

The manual may not be used for the purpose of competition and shall also not be forwarded to third 

parties. Copies for personal use are allowed. All contents, technical data and illustrations contained in 

the manual are based on information available at the time of publication.

1.1  Explanation of symbols

this symbol indicates safety information. 

It is essential that all safety information is strictly observed. Failure to comply with this  

information can lead to personal injuries or damage to property. E+E Elektronik®  

assumes no liability if this happens.

this symbol indicates instructions. 

The instructions shall be observed in order to reach optimal performance of the device.

1.2  Safety instructions

1.2.1.  General

Excessive mechanical stress and inappropriate usage may lead to failures and shall be avoided.

The sensors inside the sensing head (the tip of the sensing probe) are is an Electro Static  

Discharge (ESD) sensitive components and shall be handled as such.

Do.not.touch.the.sensing.elements.inside.the.sensing.head..The.replacement.of.the.filter.cap.shall.be.

done with utmost care not to touch the sensors. 

Installation, electrical connection, maintenance and commissioning shall be performed by  

qualified.staff.only.

The devices are designed for the operation with class III supply (EU) and class 2 supply (NA). 

The power supply must be switched off before opening the housing!

An existing Ethernet connection must be disconnected before opening the housing!

1.2.2.  Alarm module with voltages >50V (option AM2)

The optional alarm module is isolated from the low-voltage side of EE360 by a special partition; this 

must.remain.fitted.at.all.times.in.the.bottom.part.of.the.enclosure.

The EE360 enclosure must be tightly closed during operation. An open enclosure corresponds to IP00 

and exposes components carrying dangerous voltage. Any work (maintenance for instance) on the 

device.may.be.performed.by.qualified.staff.only.

1.2.3.  Integrated power supply (option AM3)

The EE360 enclosure must be tightly closed during operation. An open enclosure corresponds to IP00 

and exposes components carrying dangerous voltage. Any work (maintenance for instance) on the 

device.may.be.performed.by.qualified.staff.only.

1.3  Environmental aspects

Products from E+E Elektronik® are developed and manufactured observing of all relevant requirements 

with respect to environment protection. Please observe local regulations for the device disposal.

For disposal, the individual components of the device must be separated according to local recycling 

regulations. The electronics shall be disposed of correctly as electronics waste.

Содержание EE360

Страница 1: ...BA_EE360_e V2 0 Technical data subject to change 194475 Operation manual EE360 Moisture in Oil Transmitter...

Страница 2: ...and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the...

Страница 3: ...arm module option AM2 14 4 2 Integrated power supply 100 240 V AC option AM3 16 4 3 RS485 Module Modbus RTU option J3 16 4 4 Ethernet Module Modbus TCP option J4 17 4 4 1 Available TCP and UDP ports 1...

Страница 4: ...ng head The replacement of the filter cap shall be done with utmost care not to touch the sensors Installation electrical connection maintenance and commissioning shall be performed by qualified staff...

Страница 5: ...n the Modbus RTU interface An optional relays module can be used for alarms and process control The state of the art TFT colour display can show all measurands simultaneously and offers extensive erro...

Страница 6: ...activity aw is the relative measure for moisture in oil It represents the ratio between the actual amount of dissolved water and the maximum possible amount of dissolved water in the oil at a certain...

Страница 7: ...2 mm 0 2 not included in the scope of supply 54 2 13 104 4 09 71 2 80 5 1 0 20 166 6 54 180 7 09 153 6 02 54 2 13 166 6 54 5 1 0 20 7 0 28 Fig 4 Dimensions and mounting template of polycarbonate enclo...

Страница 8: ...to the mounting template below Mount the bottom part of the enclosure with 4 screws screw diameter 4 2 mm 0 2 not included in the scope of supply Stainless steel enclosure 75 2 95 110 4 33 73 2 87 5...

Страница 9: ...nd scaling are freely selectable All settings can be performed via display and push buttons or using the EE PCS Product Configuration Software Error indication The analogue outputs feature an error in...

Страница 10: ...g PE n c phase L1 phase L1 n c Power supply 100 240 V AC for polycarbonate enclosure Power supply 100 240 V AC for stainless steel enclosure 2 1 4 3 5 RS485 B D GND RS485 shielded RS485 A D n c Modbus...

Страница 11: ...ss 3 3 1 Installation of the probe directly in the process see Fig 10 Installation of the EE360 probe directly into the process For direct probe installation into a duct stop valves should be provided...

Страница 12: ...may deviate by max 40 C 72 F from the regular tempera ture during normal operation Installation of the probe see Fig 11 Installation of the probe by utilizing the ball valve set Install the probe into...

Страница 13: ...ly expulsed when releasing the lock not May sure nut to bend and damage the probe cable Loosen slowly the lock nut with a spanner spanner width 24 only till the probe is pushed out by the overpressure...

Страница 14: ...C 2 NO NC C Standard Inverted REL 1 REL 2 RELAY 2 RELAY 1 PROTECTIVE CONDUCTOR TERMINAL Maximum wire cross section 1 5 mm Fig 12 Alarm module option AM2 The metal housing must be grounded during opera...

Страница 15: ...Configuration Software or using the display and the push buttons Menu Alarm number Alarm output 1 Alarm 2 Alarm 1 Mode Switch hystersis Measurand Switching point Switching point Hysteresis Standard I...

Страница 16: ...st excess current and short circuit must be designated to a wire cross section of 0 8 mm AWG 18 6A fuse The metal housing must be grounded during operation National regulations for installation must b...

Страница 17: ...Ethernet module features power over Ethernet PoE and RJ45 connector with IP65 protection class The Ethernet connection must be disconnected before opening the housing Due to strong self heating effect...

Страница 18: ...otocol format as payload data See Ethernet communication protocol which can be found on https www epluse com en ser vice support download center 4 4 2 IPv4 Settings Factory setting IP Address 192 168...

Страница 19: ...0 22 or higher Firmware for Display V 1 0 9 or higher Order code Ethernet Module for retrofitting plastic enclosure HA010606 for remote probe type T5 T10 HA010607 for duct mounting type T2 4 5 Modbus...

Страница 20: ...ic unit 1 2 Barcode for probe serial number Fig 19 Pluggable probe Probe replacement procedure Plug off the damaged probe Plug on the new probe Update the 7 parameters with EE PCS Product Configuratio...

Страница 21: ...over and plug in the USB cable PCB LED 1 LED 2 LED 4 LED 3 Current output Error LED error LED on error category 1 error LED flashes error category 2 Voltage output power supply main board Fig 21 Statu...

Страница 22: ...the user from 1 second to 12 hours The data logger setup can be performed via display and push buttons Menu Sampling rate Data logging OFF 1 s 10 s 1 min 5 min 15 min 30 min 1 h 5 h 12 h The data logg...

Страница 23: ...g 23 Data logging Each point in the graph represents a logged value The points are connected by a linear interpo lation One point The data logging continues even when the data memory is full new data...

Страница 24: ...scaling error indication relay configuration mode set points state adjustment 1 and 2 point humidity temperature adjustment reset to factory adjustment calibration status configuration of Modbus data...

Страница 25: ...e 9 x 10 x not available with 0 1 V output Tab 2 Overview of error codes Error category 1 non critical error can be solved by the user The display blinks and the buzzer beeps every 10 seconds The red...

Страница 26: ...n the lock of the filter cap Remove the filter cap from the tip of the probe with utmost care not to touch the sensors Solder off the damaged humidity sensing element Shorten the leads of the replacem...

Страница 27: ...an be performed in the E E Elektronik calibration laboratory For infor mation on the E E capabilities in ISO or accredited calibration please see www eplusecal com Calibration and adjustment by the us...

Страница 28: ...TMENT NEW ADJUSTMENT Fig 28 Offset adjustment measurement value 1 Measured value Displayed value reference value 1 ADJUSTMENT POINT LOW ADJUSTMENT POINT HIGH reference point LOW factory setting 30 RH...

Страница 29: ...be material Stainless steel 1 4404 AISI 316L Enclosure material for plastic enclosure Polycarbonate UL94 V0 approved for metal enclosure Stainless steel 1 4404 AISI 316L Protection class IP65 Cable gl...

Страница 30: ...for retrofitting plastic housing HA010606 for remote probe type T5 T10 HA010607 for duct mounting type T2 1 Only for devices with PC4 option 2 2 pieces necessary per device 3 For upgrade to Modbus RTU...

Страница 31: ...settings Status Menu only available with connected modules Detailed information Menu Sampling rate Data logging Axis 1 Measurand Axis 2 Measurand Scale maximum value value value value Scale minimum OF...

Страница 32: ...ature C Temperature F Temperature K Mode Current Voltage ON Output number Output 1 Output 2 OFF Measurand min max value OK value Cancel Analog range Scaling Error indication OFF Water activity aw Wate...

Страница 33: ...asuring system Working pressure Parameters Humidity coefficient Resistance R0 Capacity C76 Capacity gain Capacity offset Temperature coefficient TK Resistance Offset Daylight saving time UTC settings...

Страница 34: ...point Switching point Hysteresis Standard Inverted Normal state Switch window Measurand Switching point low Switching point Hysteresis Switching point high Switching point Hysteresis Standard Inverted...

Страница 35: ...35 Operating instructions for EE360 Moisture in Oil...

Страница 36: ...cn BEIJING Tel 86 10 84992361 info epluse cn SHANGHAI Tel 86 21 61176129 info epluse cn E E GERMANY www epluse de Tel 49 6172 13881 0 info epluse de E E FRANCE www epluse fr Tel 33 4 7472 35 82 info...

Отзывы: