19
Operating instructions for EE360 Moisture in Oil
IP address
Subnet mask
Type
shows the status of the DHCP setting
value
value
value
value
Gateway
DNS server
Menu
IP settings
IPv6 is not supported
Supported Ethernet standard: 802.3i/u/x and af
4.4.3
Retrofit with Ethernet Module
The.EE360.can.be.retrofitted.with.an.Ethernet.Module..Before.retrofitting.please.make.sure,.that.the.
Firmware.on.your.device.is.updated.to.the.latest.version.by.using.the.Configuration.Software.EE-PCS.
•.
Firmware for Transmitter: V 1.0.22 or higher
•.
Firmware for Display: V 1.0.9 or higher
Order code:
Ethernet-Module for retrofitting plastic enclosure
HA010606 for remote probe type t5, t10
HA010607 for duct mounting type t2
4.5 Modbus Map
Register
[DEC]
Protocol
address [HEX]
Measured value
Unit
type
Read registers: function code 0x03 / 0x04
31021
3FC
Relative humidity
1)
%
31003
3EA
Temperature
°C
32-bit float
31005
3EC
Temperature
°F
32-bit float
31009
3F0
Temperature
K
32-bit float
31135
46E
Water activity
aw
32-bit float
31141
474
Water content
x
32-bit float
Write registers: function code 0x06 for 16-bit and 0x10 (decimal: 16) for 32-bit
0001
0
Slave-ID
/
16-bit integer
5001
1388
Air pressure
mbar
32-bit float
1) Use for adjustment and calibration.