20
Operating instructions for EE360 Moisture in Oil Transmitter
4.6 Pluggable probe (option PC4)
EE360 transmitters are optionally available with pluggable sensing probe, which is attached to the
EE360 enclosure by a push-pull plug. If the probe or the probe cable gets damaged it is possible to
easily replace the probe without humidity and temperature adjustment. The replacement probe (see
order information below) is supplied with a set of 7 individual parameters.
If you use two or more devices at the same time, make sure not to mix up the sensors! You can read
out the serial number of the connected sensor by scanning the barcode on the basic unit.
1.
2.
Barcode for probe serial number
Fig. 19
Pluggable probe
Probe replacement procedure
•.
Plug off the damaged probe
•.
Plug on the new probe
•. Update.the.7.parameters.with.EE-PCS.Product.Configuration.Software.or.via.display.and.push.but
-
tons.
A
B
Working pressure
Parameters
Humidity
coefficient
Resistance R0
Capacity C76
Capacity gain
Capacity offset
Temperature
coefficient.TK
Resistance Offset
Menu
Device
settings
Probe replace-
ment
Oil parameter
note:
When replacing the probe, the factory calibration loses its validity.
A calibration to verify accuracy is recommended but not mandatory.
order information for replacement probe
PE360
Cable length
2 m (6.6 ft)
K2
5 m (16.4 ft)
K5
10 m (32.8 ft)
K10
Probe length
100 mm (3.94“)
L100
200 mm (7.87“)
L200
Process conn.
1/2“ ISO thread
PA23
1/2“ NPT thread
PA25
Tab. 1
Ordering code for replacement probe