BlueWave® MX-Series 1 Channel Curing Systems User Guide
32
Figure 20. Information Button
Figure 21. Information Screen
Troubleshooting & Maintenance
Measuring/Checking Intensity
ACCU-CAL
™
50-LED radiometers are designed to measure intensity levels up to
40,000 mW/cm
2
, so they can easily handle
BlueWave
MX-Series 1CH systems. It is also
capable of exposure to temperatures up to 100°C. As with any energy source, extended
exposures will cause a thermal rise of the sensor assembly, so exposure times at higher
intensities (>20 W/cm
2
) should be limited to less than 10 seconds.
Product Cleaning
Product cleaning is limited to wiping the product with a damp cloth. Do not soak.
Isopropanol Alcohol or household cleaners may be used for cleaning the product.
Alarms
When the system malfunctions, a red alarm icon will appear, along with an audible
beep, on the Production or Admin Screen signaling that there is a problem. To check the