Administration/Production Mode Toggle
—
Switches between administration
and production modes. A password will be required to enter administrator mode. Refer
to User Settings/Boot Modes on page 30 for more information.
The production screen layout is identical to the administrator screen. The time, power
and mode settings reflect the administrator settings and cannot be altered. User
operational controls are limited to “Run/Stop”, “Volume” and “Brightness” (see Settings
screen). Program(s) names will be displayed along with the Emitter type but the ability
to save and load Program(s) will be disabled. The default password is 1234.
Settings Button
—
Loads the settings screen where the volume, language, screen
brightness, and other user settings can be adjusted.
Refer to “
” on page
24 for more information.
Irradiating Time
—
The Irradiating time is displayed in this field.
When in timed mode, time is entered by touching the display field which in turn brings
up a numerical keypad for entry. The entry range is 0.1
–
999 seconds. From 0.1
–
30
seconds the adjustment is made in 0.1 second increments and from 30
–
999 seconds
the adjustment is made in 1 second increments.
When in manual mode, time is displayed and entry is not available.
Power Button
—
The Irradiating power level is displayed in this field. The power level
is entered by touching the field which in turn brings up a numerical keypad for entry.
The power range is 10 to 100% in increments of 1%.
Run/Stop Cycle Button
—
Initiates/terminates an exposure. The button also functions
as an indicator which informs the user that the irradiator is emitting curing energy.
Operation Mode
—
Select from timer or manual mode. For more information on
operating modes, refer to “Modes of Operation”
on page 19.
Alarms Icon
—
Alerts the user that there is an error. Refer to page 32 for more
information.
Emitter Type
—
Identifies the emitter model connected to the controller and the
wavelength that it provides. Emitters are available in two models, the MX-150 spot and
the MX-250 flood. Each model is also available in three wavelengths: RediCure® (365
nm), PrimeCure® (385 nm), and VisiCure® (405 nm).
Program Button
—