![Dvorak SPIDER ILD01 G Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/dvorak/spider-ild01-g/spider-ild01-g_service-manual_2544108020.webp)
20
/
66
SM-1 Ev6/25062014
6.4 Exchange of the hydraulic pump
-
Suck out oil from the hydraulic tank
Dismount:
-
Stabilizer on the left side for better access to the pump
-
Remove the ventilator grating
-
Ventilator
-
Ventilator cover (3 screws of the pump head)
-
Picture 31
-
Tighten the screws back in the pump head
-
Hydraulic tube A - short (nuts 2 and 3 simultaneously, then 4) (Picture 32, 35)
-
Hydraulic tube B - long (loosen nut 5, then simultaneously 6 and 7) (Picture 32, 35)
-
Hydraulic hoses
-
ATTENTION: The oil must not get in the case and clutch!
-
Operating rod of the pump
Picture 32
Picture 33
1 – grating screw
2 – ventilator screw
Tube B
Tube A
Hydraulic hoses
1
1
1
2