background image

28 

1018100100L02  1907V005 

13  Information about EMC in accordance with EN 60601-1-2

13.1 General notes

The information in this leaflet includes excerpts from the relevant European standards for electrical, 
medical devices. Comply with the information provided in this leaflet when you install and combine in-
dividual Dürr Technik units with products from other manufacturers. If you are uncertain about any de-
tails, please refer to the complete standard. 

13.2 Abbreviations

EMC

Electromagnetic compatibility

HF

High frequency

U

T

Rated voltage of the device (supply voltage)

V

1

, V

2

Compliance level for the test in acc. with IEC 61000-4-6

E

1

Compliance level for the test in acc. with IEC61000-4-3

P

Rated power of the transmitter in Watts (W) in accordance with the specifications of the 
transmitter manufacturer

d

Recommended safety distance in metres (m)

13.3 Guidelines and manufacturer's information

Resistance to electromagnetic interference (immunity) for all devices and systems

The unit is designed for use in the electromagnetic environments specified below. The customer or 
operator of the unit should ensure that it is operated in such environments.

Interference immu-
nity tests

IEC 60601 - test 
level

Compliance level

Electromagnetic environment 
- guidelines

Electrostatic dis-
charge (ESD) in ac-
cordance with 
IEC 61000-4-2

±

8 kV contact dis-

charge

±

15 kV air discharge

±

8 kV contact dis-

charge 

±

15 kV air discharge

Floors should be made of wood 
or concrete, or covered with ce-
ramic tiles. If the floor is covered 
by synthetic material, then the 
relative humidity must be at 
least 30%.

Fast transient electri-
cal disturbance vari-
ables/burst in ac-
cordance with 
IEC 61000-4-4

±

2 kV for power 

cords 

±

1 kV for input and 

output cables

±

2 kV for power 

cords 

±

1 kV for input and 

output cables

The quality of the supply voltage 
should correspond to a typical 
commercial or hospital environ-
ment.

Voltage surge in ac-
cordance with 
IEC 61000-4-5

±

1 kV voltage outer 

conductor- 
outer conductor 

±

2 kV voltage outer 

conductor earth

±

1 kV push-pull volt-

age

±

2 kV common 

mode voltage

The quality of the supply voltage 
should correspond to a typical 
commercial or hospital environ-
ment.

Appendix

EN

Содержание SICOLAB med

Страница 1: ...SICOLAB med Installation and operating instructions 1018100100L02 1907V005 EN 0297...

Страница 2: ......

Страница 3: ...2 28 13 1 General notes 28 13 2 Abbreviations 28 13 3 Guidelines and manufacturer s in formation 28 EN Contents Important information 1 About this document 3 1 1 Warnings and symbols 3 1 2 Copyright...

Страница 4: ...2 1018100100L02 1907V005 Contents Contact 14 Addresses 32 14 1 Returns Repairs 32 14 2 To order spare parts 32 14 3 Service 32 14 4 Addresses worldwide 32 EN...

Страница 5: ...hey correspond to the relevant version of the unit and the status of technology valid at the time of its market launch In the event that the instructions and notes in these installation and operating...

Страница 6: ...nufacturer ac cepts no liability for damage resulting from such use In such cases the user operator will bear the sole risk WARNING Danger of explosion from the ignition of flammable materials i i Do...

Страница 7: ...ccessories and special accessories that are specified or approved by D rr Tech nik i i Only use original working and spare parts i i Ensure that the unit operator has access to the Installation and Op...

Страница 8: ...re hoses must be vented before they are stored or transported i i Protect the unit from moisture during transportation i i Always transport the unit in an upright position D rr Technik accepts no liab...

Страница 9: ...ry possible variations due to country spe cific requirements and or import regulations Product name Article number SICOLAB med 1018xxxxxx Power cord Installation and Operating Instructions 1018100100...

Страница 10: ...ow depends on the ambient pressure The actual minimum flow decreases at the ratio of ambient pressure 1013 mbar Peak flow 180 L min for max 0 8 sec Dew point reduction at operating pres sure 5 C at am...

Страница 11: ...Electrical fre quency Hz Nominal current A Output P1 kW Fuses IEC 127 2 V Units without transformer 220 240 50 2 2 T4H250V Units with transformer 220 230 60 2 4 T4H250V 115 120 50 or 60 4 7 T8H250V 12...

Страница 12: ...f the following pertinent directives Simple pressure vessel directive 2014 29 EU RoHS directive 2011 65 EU The company D rr Technik GmbH Co KG bears the sole responsibility for issuing the declaration...

Страница 13: ...d in the scope of delivery must be connected to the compressed air supply 10 of the unit and must be connected to the ventila tor If the unit is used in standby mode an addi tional pressure hose with...

Страница 14: ...the ventilator 11 Connector for central compressed air supply in standby mode 20 Air intake filter 21 Regulating valve 22 Compressor unit 23 Cooler 24 Sub micro filter 0 3 m 25 Condensate separator 2...

Страница 15: ...by mode In standby mode the ventilator is supplied from the central compressed air supply via the connection 11 with non return valve The compressor is in standby If the pressure in the central supply...

Страница 16: ...14 1018100100L02 1907V005 Product description 5 5 Device label 1 Medium 2 Supply pressure max 6 bar 3 Compressed air input symbol EN...

Страница 17: ...1018100100L02 1907V005 15 Product description 1 Medium 2 Operating pressure max 5 bar 3 Compressed air outlet symbol EN...

Страница 18: ...the unit nom 40 Kg 88 lbs medium 0297 2017 08 23 Date of manufacture Duerr Technik GmbH Co KG Pleidelsheimer Str 30 74321 Bietigheim Bissingen Made in Germany 1018101001 DSKJ 0000 REF SN 1018101001DS...

Страница 19: ...flow in and out without any obstruction The air is filtered when it is sucked in This does not alter the composition of the air The source of the air taken in should be free of any harmful substances...

Страница 20: ...om Beh lter Werk Burgau GmbH are installed in the unit The instructions for use given below apply to the following pressure vessel type Type Pressure1 Vessel2 APP3 c4 Remark5 316033 0834100014 PS 10 b...

Страница 21: ...ration However this pressure may be temporarily exceeded by up to 10 Vibration stress that would be damaging for the pressure vessel and corrosion on the ves sel must be prevented using appropriate me...

Страница 22: ...it is secured against rolling away 9 3 Establishing the compressed air connection 10 11 10 Compressed air connection of the ventilator 9 Commissioning 9 1 Remove the packaging The unit is securely pro...

Страница 23: ...ompressed air supply e g to a wall outlet of the central compressed air system for standby mode This connection is available as a NIST connection a DISS connection or a G1 4 internal thread Compressed...

Страница 24: ...onnect the power cord plug to a properly in stalled mains socket outlet with PE conductor i i Route the power cord in such a way that it is not under any mechanical tension i i Before commissioning ve...

Страница 25: ...Off switch For this purpose a compressed air gun can be connected to the compressed air connector and can be actuated until no air is replenished any longer 10 3 Unit with castors During operation th...

Страница 26: ...ection and maintenance no later than after one year or 5000 operating hours If needed service the unit at shorter intervals Maintenance and repair work may be done only by D rr Technik or D rr Technik...

Страница 27: ...preparations based on the following agents are suitable Aldehydes quaternary ammonium compounds The following preparations on the basis of phenol containing compounds halogen releasing compounds stron...

Страница 28: ...Check the fuses on the power plug replace them as required Unit does not start Air intake filter blocked i i Contact service and insert a new air intake filter cartridge Compressor is overloaded i i...

Страница 29: ...unit from the mains supply and allow it to cool down i i Provide for unobstructed air circulation i i Reduce the ambient tempera ture Soiled filter mat i i Contact customer service Have maintenance do...

Страница 30: ...t is designed for use in the electromagnetic environments specified below The customer or operator of the unit should ensure that it is operated in such environments Interference immu nity tests IEC 6...

Страница 31: ...ds 70 UT 30 drop in UT for 25 periods 0 UT 100 drop in the UT for 5 s The quality of the supply voltage should correspond to a typical commercial or hospital environ ment If the device operator needs...

Страница 32: ...investigations on sitea b Interference is possible in the environment of units that have the following symbols Comment 1 The higher frequency range applies for 80 MHz and 800 MHz Comment 2 These guide...

Страница 33: ...150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 Table 2 Recommended safety distance bet...

Страница 34: ...g units if possible Always pack the units in a plastic bag Use recyclable packing material 14 2 To order spare parts Tel 49 0 71 42 9022 0 Fax 49 0 71 42 9022 99 E mail office duerr technik de The fol...

Страница 35: ......

Страница 36: ...D rr Technik GmbH Co KG Pleidelsheimer Strasse 30 74321 Bietigheim Bissingen Germany Fon 49 7142 90 22 0 www duerr technik com office duerr technik de...

Отзывы: