22
1018100100L02 1907V005
Assembly
NOTICE
Automatic start-up of the unit after
cooling down
i
Allow unit to cool down
The motors in the units are equipped with a
temperature switch that switches the unit off if it
overheats.
The orange warning LED "Temperature too high"
is on and a permanent acoustic signal is issued.
If this happens, you will first need to determine
the cause of the fault and then remedy it, see
"Troubleshooting".
i
Guide the power cord through the strain relief
without any mechanical stress and screw
down the strain relief.
i
Connect the power cord plug to a properly in-
stalled mains socket.outlet with PE conductor.
i
Route the power cord in such a way that it is
not under any mechanical tension.
i
Before commissioning, verify that the power
supply voltage complies with the voltage
specifications of the type plate.
DANGER
Risk of electric shock due to dam-
aged power cord or plug
Electric shocks can cause severe inju-
ries.
i
Do not start up the unit if the power
cord or plug is damaged.
i
Replace the damaged power cord.
DANGER
Electric shock due to missing protec-
tive earth
i
Connect the unit to the protective
earth (PE) connection.
9.6 Overtemperature protection
If the unit is switched off via the tempera-
ture switch, the fan of the unit will contin-
ue to run.
EN