Assembly
NOTICE
The use of a non-approved power
cord can have a negative effect on
the electrical safety and EMC.
The use of non-approved accessories,
e.g. power cord, can give rise to in-
creased electromagnetic emissions and/
or reduced electromagnetic immunity.
i
Use original accessories (power cord)
of the manufacturer exclusively.
If the power cord is missing or damaged,
please contact Dürr Technik customer
services or the distributor.
i
Plug the built-in plug (C14) of the
power cord into the IEC socket (C13 accord-
ing to IEC 60320) on the rear of the unit.
i
Tighten the strain relief anti-clockwise using
the AF 3 Allen wrench.
14
G
3
14 Strain relief
11 Connector for central compressed air sup-
ply in standby mode
There are two compressed air connectors on
the rear of the unit. The upper compressed air
connection (10) is connected to the ventilator
via a pressure hose. The compressed air con-
nection is available as as NIST connection, DISS
connection or a NW 7.2 quick coupling.
The lower connector (11) is connected to the
central compressed air supply (e.g. to a wall
outlet of the central compressed air system) for
standby mode. This connection is available as a
NIST connection, a DISS connection or a G1/4"
internal thread.
Compressed air connection for the ventilator
The compressed-air connector is set to a pres-
sure of 4.2 bar at the factory.
The pressure hose routed from the unit to the
ventilator must not be routed through cool sur-
roundings (e.g. placed on the floor). It may be
routed to be as short as possible and without
kinking. Otherwise there may be condensation
in the pressure hose.
Connector for central compressed air supply
For units used in standby mode. This
means for supply of the ventilator follow-
ing outage of the central compressed air
system.
i
Connect the flexible pressure hose to connec-
tor (11) on the rear of the unit and to the wall
outlet of the central compressed air system.
9.4 Condensate
Any condensate is evaporated by an evapora-
tion system.
9.5 Electrical installation
NOTICE
Improper use of the power cord leads
to malfunction
The enclosed power cord is intended
exclusively for use with the delivered
unit. If the power cord is used for other
units, these units need to be monitored
for proper function.
i
Use power cord for the delivered unit
exclusively.