Using the PRIORITY Button
Assigning Station Call Priorities
To assign a call priority to a station, your base master must cover the station to be
changed, either by capturing, swinging, sharing, or permanent coverage.
—IMPORTANT UL NOTICES—
A patient normal call is canceled when it is answered at the master station, or when the
CANCEL button is pressed on the station that placed the call. If your system includes
wireless phones, this type of call will also be canceled if it is answered by a wireless
telephone. BE AWARE that your ProCare 6000 System will no longer be UL Listed if a
patient call is canceled anywhere other than the master station or the station that placed
the call (i.e., if it is canceled by a wireless phone). To maintain your system’s UL Listing,
program the patient stations to place calls that are patient priority or patient emergency.
If your system includes IR staff location, a patient normal call will be canceled when staff
enter the calling room. BE AWARE that your ProCare 6000 System will no longer be UL
Listed if a patient call is canceled anywhere other than the master station or the station that
placed the call (i.e., if it is canceled by an IR staff location system). To maintain your sys-
tem’s UL Listing in this situation, program the patient stations to place calls that are
patient emergency.
1. Call the station to be assigned a different call priority. See
Placing a Call
on page 2-4.
Note:
You must call the station. If the station calls you, you cannot change the priority.
2. Press
PRIORITY
when the station number and “TALK” appear in the display.
3. Press
1
to select Pset: Bed Call = 1. The current station call priorities are available:
DEFAULT
Bed Call: NORM = 1
Bed Call: PPRI = 2
Bed Call: PEMG = 3
Review = VOL DOWN
YOUR SYSTEM PRIORITY NAMES
Bed Call: ____________________ = 1
Bed Call: ____________________ = 2
Bed Call: ____________________ = 3
Review = VOL DOWN
Note:
Your base master may also show selections beyond the three shown above. These are
used when auxiliary equipment is connected to the station. DO NOT select these
priorities here; instead, see
Assigning Station Auxiliaries or Station Prime Auxiliaries
on page 2-20 for explanations and instructions.
continued
ProCare 6000 Operation Manual
2-19
Using the PRIORITY Button