Copyright 2010 Baker Hughes Company.
54 | PACE Indicators Instruction Manual–English
Chapter
6.
Reference
6.9.2 Analog Output Option
The Analog Output option provides an analog output that is proportional to measured pressure.
Table
6
-
5: Pin Number and Function
Pin Number
Function
Pin Number
Function
1
(not used)
9
(not used)
2
(not used)
10
0
V return
3
(not used)
11
+24
V
dc output, 100
mA max
4
(not used)
12
Switch In 1
5
(not used)
13
Switch In 2
6
(not used)
14
Analog
7
(not used)
15
Analog output -
8
(not used)
Selects Analogue
Output Range
On/Off
Update rate of Analogue Output
option from the control module.
Analogue Output Bandwidth = 0.5 x Update Rate (Hz)
30V maximum with respect to
chassis.
Rated output = 24V
30V
max
To maintain PACE product safety, external circuits
connected to the instrument must meet Safety
Extra-Low Voltage (SELV) requirements.
15-way female D connector
Front view of Analogue
Option PCB connector
1
2
3
4
5
6
7
8
15
14
13
12
11
10
9
30V
max
22V to 26V
ANALOGUE 1
ANALOGUE 2
15
CONSULT HANDBOOK
analogue connections
Содержание PACE1000
Страница 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Страница 2: ......
Страница 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Страница 12: ...Copyright 2010 Baker Hughes Company 2 PACE Indicators Instruction Manual English Chapter 1 Description...
Страница 24: ...Copyright 2010 Baker Hughes Company 14 PACE Indicators Instruction Manual English Chapter 2 Installation...
Страница 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Страница 30: ...Copyright 2010 Baker Hughes Company 20 PACE Indicators Instruction Manual English Chapter 3 Operation...
Страница 70: ...Copyright 2010 Baker Hughes Company 60 PACE Indicators Instruction Manual English Appendix B Air Density...
Страница 75: ......