6.11 Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it to the nearest Druck Service Centre
Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA).
Provide the following information for a RGA or RMA:
•
Product (e.g. PACE1000).
•
Serial number.
•
Details of defect/work to be undertaken.
•
Include any error code(s) if available.
•
Operating conditions.
6.11.1 Safety Precautions
You must inform Druck if the product has been in contact with any hazardous or toxic substance.
The relevant COSHH or in the USA, MSDS, references and precautions to be taken when
handling.
6.12 Packaging Procedure
1. The instrument should be at zero/ambient pressure.
2. Switch off and isolate the electrical power supply to the instrument.
3. Shut off the pneumatic pressure and vacuum supplies to the instrument.
4. Remove the instrument from the equipment rack to access the rear panel.
5. Disconnect the power supply cable and the pneumatic supply hose assemblies.
6. Stow the power supply cable in the packaging below.
7. Remove any pressure adaptors, diffusers and restrictors.
If available, use the original packing material. When using packing materials other than the
original, do the following:
8. Fit protection to all the ports to prevent ingress of moisture and dirt.
Note:
Use the original red plastic plugs or low tack masking tape.
9. Wrap unit in polyethylene sheeting.
10. Select a double-wall cardboard container.
•
Inside dimensions must be at least 15
cm (6”) greater than the equipment
•
The carton must meet test strength requirements of ≥
125 kg (275
lbs).
11. Protect all sides with shock-absorbing material to prevent equipment movement within the
container.
12. Seal carton with approved sealing tape.
13. Mark carton “FRAGILE” on all sides, top, and bottom of shipping container.
The following conditions apply for both shipping and storage:
•
Temperature range: -20° to +70°C (-4° to +158°F).
INFORMATION
Service by unauthorized sources will affect the warranty and
may not guarantee further performance.
Содержание PACE1000
Страница 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Страница 2: ......
Страница 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Страница 12: ...Copyright 2010 Baker Hughes Company 2 PACE Indicators Instruction Manual English Chapter 1 Description...
Страница 24: ...Copyright 2010 Baker Hughes Company 14 PACE Indicators Instruction Manual English Chapter 2 Installation...
Страница 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Страница 30: ...Copyright 2010 Baker Hughes Company 20 PACE Indicators Instruction Manual English Chapter 3 Operation...
Страница 70: ...Copyright 2010 Baker Hughes Company 60 PACE Indicators Instruction Manual English Appendix B Air Density...
Страница 75: ......