Copyright 2010 Baker Hughes Company.
52 | PACE Indicators Instruction Manual–English
Chapter
6.
Reference
Note:
An English and French language file name would be: LanguageEnglish.lng and
LanguageFrench.lng. Language files named “Language.lng” or in any other format will be ignored
by PACE.
Figure
6
-
1: Language Setting
6.6.9 Restore as Shipped Settings
Restores instrument settings to factory default.
Note:
Does not affect PIN settings.
6.7 Calibration
The instrument should be returned to the manufacturer or calibration facility. Refer to
6.12, “Packaging Procedure,” on page
To find the date of the last calibration, press Measure Setup > Status > Calibration history.
The calibration menu provides facilities for programming settings for maintenance as follows:
6.7.1 Sensor Correction
Selects the range for a three-point calibration routine.
INFORMATION
A PIN protects the Calibration menu against unauthorized
use. Each instrument, on delivery, contains the factory set PIN (4321). To
continue protecting the supervisor setup menu, the PIN should be changed as
soon as possible.
Содержание PACE1000
Страница 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Страница 2: ......
Страница 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Страница 12: ...Copyright 2010 Baker Hughes Company 2 PACE Indicators Instruction Manual English Chapter 1 Description...
Страница 24: ...Copyright 2010 Baker Hughes Company 14 PACE Indicators Instruction Manual English Chapter 2 Installation...
Страница 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Страница 30: ...Copyright 2010 Baker Hughes Company 20 PACE Indicators Instruction Manual English Chapter 3 Operation...
Страница 70: ...Copyright 2010 Baker Hughes Company 60 PACE Indicators Instruction Manual English Appendix B Air Density...
Страница 75: ......