Страница 2: ......

Страница 3: ...a pressure reading in various pressure measurement units These user instructions include the operations for all DPI 705E Pressure Indicators safety instructions and the requirements for intrinsically...

Страница 4: ...rsions refer to the Quick Start and Safety Manual provided with the unit and also available for download on Druck com essential Pressure Connectors and Adaptors The DPI 705E and PM 700E are fitted wit...

Страница 5: ...uck com essential The instrument requires 4 x AA LR6 batteries fitted as shown below Use a small Posidrive screwdriver to loosen the screws in the battery compartment cover and fit as per the battery...

Страница 6: ...e internal pressure sensor the external sensor must be unplugged The external sensor may be either A pressure sensor PM 700E ranging from 25 mbar to 1 400 bar 0 36 psi to 20 000 psi full scale A tempe...

Страница 7: ...ng the Tare feature It has a secondary function as an Increment key Units Change the measurement unit of the instrument It has a secondary function as a Next key Backlight Turn the instrument backligh...

Страница 8: ...nal Indicator 5 Calibration Due Indicator 6 Remote State Indicator 7 Internal Sensor Functional Indicator 8 Lock Indicator 9 Alarm State Indicator 10 Temperature Degree Indicator 11 Resistance Ohms In...

Страница 9: ...12 3 1 Advanced Settings 13 3 1 1 Adjust Sensor Calibration 13 3 1 2 Calibrate Sensor 15 3 1 3 Set Calibration Date 17 3 1 4 Set Calibration Due Interval 17 3 1 5 Change User PIN 18 3 1 6 Units Lock...

Страница 10: ...Copyright 2019 Baker Hughes Company viii 137M7122 DPI 705E DPI 705E IS User Manual English...

Страница 11: ...ry level is displayed 3 The positive full scale of the sensor is briefly displayed 4 The number of days remaining to the next calibration due date are then briefly displayed INFORMATION If no date has...

Страница 12: ...P1 pressure port at the top of the instrument 2 The instrument will display the measured pressure using the internal sensor as shown below 3 To measure pressure via an external sensor connect an exter...

Страница 13: ...ero should be performed on a Gauge or Differential pressure instrument before measuring pressure The zero function is not available on absolute pressure sensors because absolute vacuum would have to b...

Страница 14: ...If tare is active then will be shown on the display as below 1 6 FILTER The filter functionality allows a filtered pressure reading by showing a rolling average of last 10 measurements This will give...

Страница 15: ...press including brief press of backlight key Permanently on mode 1 Press and Hold the backlight key on the instrument until the backlight turns on 2 A double beep is heard if audio is enabled 3 The b...

Страница 16: ...t the maximum pressure reading press and hold the key until a beep sound is heard if audio is enabled 5 The reading will then reset and the unit will immediately start measuring again but is still onl...

Страница 17: ...g again but is still only showing the minimum reading on the display this is effectively negative Peak Hold mode 7 Press the key to exit minimum display mode 2 3 Leak Test This feature is used to dete...

Страница 18: ...ottom of the display 7 To leave the Leak Test screen press the key 2 4 Alarm The alarm feature allows an alarm to be set based on certain conditions For example pressure value above or below user set...

Страница 19: ...Note The instrument will not trigger an alarm if the alarm state is set to OFF See Section 2 4 1 and Section 2 4 2 to set high low alarm thresholds High and low alarms are independent of each other E...

Страница 20: ...to increment the current digit value 7 Use the key to move to the next digit 8 When all digits have been processed press the key to save the value Note To set a negative value press the key until the...

Страница 21: ...dash above the arrow indicates that this is a negative value 5 To change this value press the key The current digit will flash briefly 6 Use the key to increment the current digit value 7 Use the key...

Страница 22: ...s these settings 1 Ensure the display is on the sensor reading screen 2 Press the and keys simultaneously Advanced Feature Mode Advanced Settings General Settings Enter PIN to Unlock Not Available wit...

Страница 23: ...same procedure is used for the internal and external sensor The display shows Internal Sensor when calibrating the internal sensor and External Sensor for the external sensor Note All sensors allow a...

Страница 24: ...show Advanced which indicates the Advanced Settings option is selected 4 Press the key to enter Advanced Settings 5 A 4 digit PIN is prompted for 3 1 1 1 To Enter User PIN a Press the key to incremen...

Страница 25: ...press the key to proceed to the sensor calibration screen 2 The display will show Apply 3 Apply a known pressure or resistance for RTD from a calibrated source wait for the reading to be stable 4 Onc...

Страница 26: ...being averaged to give a more stable pressure value The clock will average the readings for 5 seconds 10 Once the clock symbol stops rotating and disappears press the key 11 Enter the pressure resist...

Страница 27: ...e to Today as shown above The unit will use the current date saved in the system as the last calibration date of the sensor 9 To accept this value press the key or press the key to cancel and leave th...

Страница 28: ...above Section 3 1 1 1 6 If the PIN is correct the display will show Adjust 7 Press the key twice to navigate to the Due screen as shown below 8 Press the key to enter the calibration interval screen T...

Страница 29: ...show Adjust 7 Press the key 3 times to navigate to the PIN screen as shown below 8 Press the key to navigate to the PIN entry screen as shown above 9 Enter the new PIN value by using the TARE key to i...

Страница 30: ...iously disabled units 1 Ensure the display is on the sensor reading screen 2 Press the and keys simultaneously 3 The display will show Advanced which indicates that the settings option is selected 4 P...

Страница 31: ...3 The display will show Advanced which indicates that the settings option is selected 4 Press the key to enter Advanced Settings 5 Enter PIN The steps to enter PIN are indicated in above Section 3 1...

Страница 32: ...the instrument to be reset to factory settings To perform a factory reset 1 Ensure the display is on the sensor reading screen 2 Press the and keys simultaneously 3 The display will show Advanced whic...

Страница 33: ...he key to return to the Advanced Settings screen 12 Press the key again to return to the sensor reading screen 3 2 General Settings The following general settings are available on the DPI 705E Setting...

Страница 34: ...as seen in above example 2000 mbar 7 To exit from this screen either press the key again or press the key 8 Press the key to return to the sensor reading screen 3 2 2 Calibration Due Date The calibrat...

Страница 35: ...ration as shown below e g 154 days Note If an external sensor is connected then the display will show External Sensor and the data for that sensor will be displayed The calibration information includi...

Страница 36: ...and return to the previous screen or press the key to return without saving 9 Press the key to return to the sensor reading screen 3 2 4 Auto shutdown This feature will allow control of the auto shut...

Страница 37: ...key 8 To accept the value and return to the previous screen press the key 9 To leave the screen without saving press the key 10 Press the key to return to the sensor reading screen 3 2 5 Battery Type...

Страница 38: ...ey to enter the battery type setting option 6 The default battery type is Alkaline displayed as Battery Alkaline shown below 7 To change the battery type press the key This will show the next battery...

Страница 39: ...w Advanced which indicates that the unit is now in the Settings option 4 Press the key 6 times to navigate to the Date screen as shown below 5 Press the key to enter the date setup 6 The year value is...

Страница 40: ...nt flashing digit and the key to move to the next digit 11 Press the key to save the changes and return to the settings menu or press the key to return without saving 12 Press the key to return to the...

Страница 41: ...and the key to move to the next digit 8 Press the key to save the changes and return to the previous screen or press the key to return without saving 9 Press the key to return to the sensor reading sc...

Страница 42: ...s option 4 Press the key 8 times to navigate to the Scale screen as shown below 5 Press the key to enter the custom unit setup 6 The current unit will be shown by default along with a down arrow e g 1...

Страница 43: ...ther units To verify this press the key and navigate to other units The scale factor values updates For example after setting a scale factor of 3 000 for mbar a scale factor of 0 0030 for bar is shown...

Страница 44: ...d which indicates that the unit is now in the Settings option 4 Press the key 9 times to navigate to the Software screen as shown below 5 Press the key to enter the software option 6 The current softw...

Страница 45: ...APT QF Quick fit hose pressure adaptor IOHOSE NP1 Nylon hose 1 m 3 3 ft length 20 bar 300 psi MWP IOHOSE NP2 Nylon hose 2 m 6 6 ft length 20 bar 300 psi MWP IO620 HOSE P1 Pneumatic Hose 1 m 3 3 ft len...

Страница 46: ...rm valid calibration or press the Back key to exit the function Err 05 Invalid calibration point applied during user calibration Press the OK key and enter valid data range or press the Back key to ex...

Страница 47: ...Applied pressure is outside range of sensor common with 1 bar 14 5 psi range absolute sensors Use within sensor limits Full scale of sensor is displayed on power up and etched on sensor G1 8 connecto...

Страница 48: ...tact the Service Department to obtain a Return Goods Material Authorization RGA or RMA Provide the following information for a RGA or RMA Product e g DPI 705E or RTD INTERFACE IS Serial number Details...

Страница 49: ......

Страница 50: ...akerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken...