5. Error Codes & Diagnostics
5.1 Error Codes
Error Code
Meaning
Solution/Action
Err.01
Entered PIN is not correct
Press the
(OK) key and enter correct PIN,
or press the
(Back) key to exit the function.
Err.02
PIN mismatch during PIN setting –
second PIN does not match first PIN
Press the
(OK) key and enter correct PIN,
or press the
(Back) key to exit the function.
Err.03
Invalid values entered (e.g calibration
attempt beyond sensor range)
Press the
(OK) key and use valid
calibration points, or press the
(Back) key
to exit the function.
Err.04
User entered value is not valid (e.g
date dd.mm = 31.02, or a user scaling
factor = 0.0)
Press the
(OK) key and perform valid
calibration, or press the
(Back) key to exit
the function.
Err.05
Invalid calibration point applied during
user calibration
Press the
(OK) key and enter valid data
range, or press the
(Back) key to exit the
function.
Err.06
User attempt to zero a sensor failed
Press the
(OK) key and enter valid data
range, or press the
(Back) key to exit the
function.
Err.07
Calibration attempted without valid
date set
Set the date/time (this is internally battery-
backed)
Err.21
Err.23
Err.24
Remote Sensor Error (pressure or
RTD sensor)
Disconnect and reconnect remote sensor – if
error persists then sensor is either incompatible
type with DPI
705E / DPI
705
-
IS or is faulty.
Err.40-63
Internal error
Internal fault - return unit to service center.
Err.64
User calibration rejected
Press the
(OK) key or the
(Back)
key to return to measurement screen.
Err.80
Critical Software Fault
Power off and back on. If problem persists return
unit to service center.