24 | 137M7122 DPI 705E / DPI 705E IS User Manual–English
3.2.1 Full-scale Value
This setting allows the full-scale value of the pressure sensor to be viewed. This is a display only
setting, the value cannot be changed.
To see the full-scale value:
1. Ensure the display is on the sensor reading screen.
2. Press the
,
and
keys simultaneously.
3. The display will show “
” (Advanced) which indicates that the unit is now in the
Settings
option.
4. Press the
key to navigate to the “ ” (full-scale) screen as shown below.
5. Press the
key to view the full-scale value.
6. The full-scale value will be shown as seen in above example, 2000
mbar.
7. To exit from this screen either press the
key again or press the
key.
8. Press the
key to return to the sensor reading screen.
3.2.2 Calibration Due Date
The calibration due date is the date when the sensor is next due for calibration.
To see the days until the calibration due date:
Time
Allows setting of system time
Custom Unit
Allows configuration of a custom scaling factor against
predefined units
Software Version
Displays the application software version of the
instrument
INFORMATION
To display the full-scale value, the appropriate units must be
selected. For example, if the full-scale value is 200
bar and mbar units are
selected (Full-scale 200,000
mbar), then the full-scale screen will show ‘
-----
’
because there are insufficient digits to display 200,000.
Setting
Description