Copyright 2019 Baker Hughes Company.
iv | 137M7122 DPI 705E / DPI 705E IS User Manual–English
External Sensors
The DPI
705E supports two types of external sensor, both of which connect via a cable into the
right-hand side of the unit.
The sensor must be connected via supplied connection cable (it will not engage correctly if
directly plugged in). The connectors push in (with polarity protection features aligned) then the
outer body rotates clockwise to lock the connector in position. Multiple cables may be used in
series (i.e. as extenders). Up to a maximum of 10 in series if very long cabled connections are
needed.
When an external sensor is connected it is automatically detected and has precedence over the
internal sensors. The display shows “
” (External Sensor) to show that the reading
is from the external sensor. To show a reading from the internal pressure sensor the external
sensor must be unplugged.
The external sensor may be either:
•
A pressure sensor PM
700E ranging from 25 mbar to 1,400 bar (0.36
psi to 20,000
psi)
full
-
scale.
•
A temperature sensor adaptor RTD
-
INTERFACE which supports a 4-wire PT100 probe.
If a PM
700E pressure sensor is plugged in then the pressure Full-scale will be displayed (only
at first plug-in), followed by the calibration due date in days. Normal pressure measurement
display will then start.
If an RTD
-
INTERFACE temperature sensor is plugged in then temperature/resistance full-scale
will be displayed (only at first plug-in), followed by the calibration due date in days. Normal
resistance or temperature measurement display will then start.
RTD Wiring
The DPI
705E M12 connector is to be connected to an external RTD probe as shown below.
The RTD
-
INTERFACE is optionally supplied with a field-rewirable M12 connector for customers
to connect their own wire-ended RTD's into. This is accessory part number IO-RTD-M12CON.
External sensor connector