Copyright 2019 Baker Hughes Company.
English–137M7122 DPI 705E / DPI 705E IS User Manual | 11
1. Ensure the display is on the sensor reading screen.
2. Press the
and
keys simultaneously.
3. The current state of the alarm will be displayed as a “ ” (On) or “
” (Off). Ensure the alarm
is set to ON.
4. Press the
key twice to navigate to the low alarm screen. The pressure value set for
low alarm will be displayed.
Note:
•
The down arrow on the left indicates that this is the low value.
•
The dash above the arrow indicates that this is a negative value.
5. To change this value, press the
key. The 'current digit' will flash briefly.
6. Use the
key to increment the current digit value.
7. Use the
key to move to the next digit.
8. When all digits have been processed, press the
key to save the value.
Note:
•
To set a negative value, press the
key until the up-arrow and down-arrow
symbols flash, then press the
key to toggle the negative value symbol.
•
Setting an invalid threshold will result in an error message. For example, Error.03 which
means the value is out of bounds. If this happens press the
key and enter the
value again.
9. Press the
key to go to the main screen.