The same procedure is used for the internal and external sensor. The display shows
“
” (Internal Sensor) when setting the calibration due interval of the internal sensor
and “
” (External Sensor) for the external sensor.
1. Ensure the display is on the sensor reading screen.
2. Press the
,
and
keys simultaneously.
3. The display will show “
” (Advanced) which indicates the Settings option is selected.
4. Press the
key to enter
Advanced Settings.
5. Enter PIN. The steps to enter PIN are indicated in above Section
6. If the PIN is correct, the display will show “
” (Adjust).
7. Press the
key twice to navigate to the “
” (Due) screen as shown below.
8. Press the
key to enter the calibration interval screen. This will show the current set
interval which set to 365 (days) by default. If this has changed previously, it will show the last
saved interval value.
9. Enter the interval value in days (between 1 and 540), using the
key to increment the
value, and the
key to move to the next digit. Repeat this for all digits.
10. Press the
key to save the entered value.
11. Press the
key to return to the
Advanced Settings
screen.
12. Press the
key again to return to the sensor reading screen.
3.1.5 Change User PIN
This feature allows the default user PIN to be changed.
To change user PIN: