HPG 13150 400 V (A)
Troubleshooting and Customer Support
6-11
HPG™ 13150 400 V RF Power Generator
Prerelease version
E30 Plate Voltage Too
Low
Plate voltage is below lower limit.
Cause: Mains voltage too low for selected transformer
tap.
Solution: Switch HPG generator off and check mains
transformer tap settings to make sure that they match the
input voltage.
Note:
Changing the taps is a technical procedure that can
only be done by properly trained personnel. For
more information, see the HPG generator service
manual.
The
ON/OFF
power switch does not completely
disconnect the
Mains
. You must have an external switch
installed to completely disconnect the
Mains
.
E31 Plate Voltage Too
High
Plate voltage was above upper limit.
• HV is switched off.
Cause: Mains voltage too high for selected transformer
tap.
Solution: Switch HPG generator off and check mains
transformer tap settings to make sure that they match the
input voltage.
Note:
Changing the taps is a technical procedure that can
only be done by properly trained personnel. For
more information, see the HPG generator service
manual.
The
ON/OFF
power switch does not completely
disconnect the
Mains
. You must have an external switch
installed to completely disconnect the
Mains
.
Table 6-1. Error description and resolution (Continued)
Error Message
Description and Resolution
Содержание 13150 400 V
Страница 2: ......
Страница 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Страница 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...