6-10
Troubleshooting and Customer Support
HPG 13150 400 V (A)
Dressler
®
HF-Technik GmbH
Prerelease version
E21 Cavity Humidity Too
High
Power section/cavity humidity is too high.
• HV is switched off.
• HPG will reinitialize after error condition has gone.
Cause: Humidity sensor at cavity reports too high
humidity.
Solution:
1. Switch off generator and disconnect the mains.
The
ON/OFF
power switch does not completely
disconnect the
Mains
. You must have an external switch
installed to completely disconnect the
Mains
.
2. Turn off the water source.
3. Check the water connections for leaks.
4. Check water temperature (low water temperature
causes condensation).
5. Dry any condensation in the generator before
reconnecting mains.
E22 Cooling Water
Conductance Too High
Grid current is negative.
Cause: The cooling water is too conductive.
Solution: Make sure that the cooling water is within
specifications (see
).
Table 6-1. Error description and resolution (Continued)
Error Message
Description and Resolution
Содержание 13150 400 V
Страница 2: ......
Страница 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Страница 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...