22
ROOTS
®
Model 5 Prover Software Manual
b.
Plug the Optical Scanner cable into the
Field Meter Junction Box at the
connection labeled I.D. PULSER.
c.
Focus the light emitted by the Optical
Scanner so that the dot displayed on
the dial is of uniform brightness and
that it does not cross more than one
black/white boundary.
d.
Adjust the Optical Scanner amplifier to
ensure that the Amplifier blinks for
each transition from white-to black or
black-to-white. To properly adjust the
amplifier on the Optical Scanner, see
the Installation and Operation manual
for the scanner. If using a different
Optical Scanner, follow the adjustment
procedures for that specific model
per the operating instructions of
that scanner’s manual.
18
. If using a Manual Stop/Start Cable and
Button as a test control mode, perform this
step then skip to step 21.
a
. Plug the Manual Start/Stop Cable
into the Field Meter Junction Box at
the connection labeled I.D. Pulser.
19.
If using a Field Counter Pulser as a test
control mode, perform this step then
skip to step 21.
a.
Attach the Field Counter Pulser to
the end of the Series 3 Lexan (unit
must be pulser ready). Plug the ID
pulser cable into the Field Meter
Junction Box labeled I.D. Pulser.
Connect the other end to the Field
Counter Pulser.
Set up the Field Counter Pulser by using Instrument Drive Pulser as the test control mode
while configuring a test.
20.
If using a SmartProve
™
Interface Cable as a test control mode, refer to SmartProve
Installation Manual
21.
Hardware set-up is now complete. Push the Controller’s power switch to the ON position
and verify that the red power indicator light (located next to the power switch on top of the
Controller) is illuminated. Proceed to the section entitled “Configuring a Meter Test” for
the software configuration and test execution.
Figure 5: Optional Scanner Relative Positioning and
Alignment Series B3 Accessory Unit.
Figure 10: Manual stop/start cable set-up.