9
Also the FIFO communication speed adjustments should be on the lowest settings. This can be reached by
going to Control Panel:Ports:Com Port Settings.
2.4 Verifying the Proper Files Are Loaded
For each Master Meter available for use with the type of ROOTS Model 5 Prover that is currently being used,
read the identification tag attached to the Master Meter(s). Verify that the serial number for each Master Meter
matches the serial number displayed in the initial start up screen of the ROOTS Model 5 Prover software. If one
of the serial numbers does not match the appropriate serial number on the screen, the original factory diskette
(or a copy of it) must be obtained in order to replace the current factory preset file(s). These file names are
P2M.SET, P5M.SET, P10M.SET, P20M.SET and/or P80M.SET and must be copied into the same directory as
the current Prover Software is located, which should be the \\Program Files\Dresser Inc\Model 5 Transfer
Prover\Config Files\subdirectory. The Startup disk should have the same serial number(s) written on the disk
label to match the serial number(s) of the Master Meter(s) on the Prover.
If the installation is a software upgrade and the disk label does not have serial numbers on the disk then the
preset files must be located and put in the same location as the new program files. During the installation
process the location of the new program will be evident. Following is a list of the configuration files that must
be copied to the new location; they are based on the style of Prover that you are using.
Common files for all prover sizes:
• calidate.cfg, calibr.cfg, and MasterMeters. key
and specific files depending on prover size:
• 2M/10M Model 5 Prover: P2M.SET, P10M.SET
• 10M only Model 5 Prover: P10M.SET
• 20M only Model 5 Prover: P20M.SET
• 5M/20M only Model 5 Prover: P5M.SET, P20M.SET
• 10M/80M Model 5 Prover: P10M.SET, P80M.SET
• 80M only Model 5 Prover: P80M.SET