User Manual of CIC-D100 Ion Chromatograph
69
:
CIC-D100 instrument does not require pressure-protected eluent bottle.
However, if the eluent is degassed manually or easily contaminated, it is recommended to
equip the eluent bottle with nitrogen protection.
B.10 Flushing pump
Before flushing, ensure that the eluent bottle is full of eluent; the eluent bottle cap is
tightened; the eluent pipeline is connected and the waste liquid pipe is inserted into the waste
liquid bottle.
B.10.1 Flushing pump head
:
Caution: When the instrument is used for the first time, for a long time, or
when it is not normally used, the pump head should be rinsed clean to avoid residual
crystals in the pump head damaging the pump head.
The flushing steps are as follows:
1. Connect the back flushing pipeline of the pump body;
2. Use a 10 mL injector to suck up ultrapure water and connect it to one end of the flushing
pipeline, and insert the outlet pipeline into the waste liquid bottle;
3. Slowly push the injector to clean the pump head. This operation can be repeated several
times.
B.10.2 Flushing eluent pipeline
:
When installing and replacing the eluent for the first time or when there is no
liquid in the eluent pipeline, the eluent pipeline should be flushed.
The flushing steps are as follows:
1. Insert a 10 mL injector into the curved needle beside the exhaust valve;
Caution
Caution
Caution