User Manual of CIC-D100 Ion Chromatograph
49
2. Excessive indoor temperature leads to incomplete degassing of ultra-pure water
Solution: adopt on-line degasser.
3. The filter head of infusion pump is blocked
Solution: Remove the filter head and put it into 1: 1 pure water/nitric acid solution or
absolute ethyl alcohol for ultrasonic cleaning.
5.14 Handling of abnormal instrument control
1. Device type query failed
Reason: After the software connects the instrument successfully, the type of the instrument
will be inquired first. If no response is received or the returned information is incorrect, it will be
displayed in the window.
Solution: Please make sure the instrument is turned on normally.
2. The reverse control cannot control the instrument.
Solution: Restart the instrument or reverse control software.