User Manual of CIC-D100 Ion Chromatograph
33
Figure 3-20 Interface button
a Sequence start button
This button is to realize one-key sequence startup and shutdown of the equipment. The
startup sequence is executed as set in the main menu [Function]-> [Startup Setting]. This button
is a ping-pong button. Press it again to shut down.
b Button of starting the collection
This button is to start chromatograph data collection and spectrogram data transmission. If
the option of "Start Collection Automatically after Startup" is selected on the spectrogram
software setting page in [Parameter Setting] under the system menu, this button will be pressed
automatically when the program is started. Otherwise, it needs to be started manually.
:
The injection signal is only responded when the collection is started.
c Baseline zero button
Send an instruction to control the spectrogram software to return to zero once.
(2) Plunger pump area
Figure 3-21 Plunger pump area
a Setting button
This button sets the plunger pump flow rate. The setting value is the value in the left flow
input box.
b Query button
Query the plunger pump flow setting value. The query results are displayed in the left flow
input box.
c Start button
This button is to start the plunger pump. It is also a ping-pong button. Press it again to close
the plunger pump. After the plunger pump is started, the real-time pressure data will be displayed
in the pressure display frame to the left of the button.
Caution