User Manual of CIC-D100 Ion Chromatograph
57
rear end of the instrument;
2. Use a flat screwdriver or other tools to gently remove the fuse fixing clip from the
instrument;
3. Remove the fuse from the fixing clip and observe whether the fuse in the glass shell is
broken in light.
4. If the fuse is broken, please replace the fuse of the same type and specification;
:
Please do not replace other types of fuses at will. If you cannot confirm the fuse
type, please contact the instrument manufacturer.
5. In addition to the same specification and type, the new fuses in the glass shell should be
in good condition and its metal shells at both ends should be free from rust.
6. During installation, ensure that the fuse is pushed into the original position after it is in
the middle of the fixing clip;
7. Connect the "pin" power plug of the instrument;
8. Turn on the instrument power switch;
9. Use the software or panel to see if the instrument returns to normal operation.
Figure 6-6 Replacement of the fuse
Through the above operations, if the instrument returns to normal operation, the fuse
replacement is completed. If there is no normal operation, please check other possible causes of
abnormal operation of the instrument.