14
Instructions for use Flowmeter
English / English US
English / English US
Fault-Cause-Remedy
Maintenance
This chapter describes the maintenance measures
required to maintain the proper functioning of the
medical device. Maintenance measures must be
performed by the personnel responsible.
Definition of maintenance concepts
Inspection
Perform inspections at regular intervals and observe
the following specifications.
.
Performing inspection
1
Check completeness and readability of labels
2
Visual inspection of
– connectors
– gas type code conformity
– the sealing ring at the gas outlet
– the CS connection hose, if available
– the pressure dome and the respective seal-
ing ring for cracks and contamination
3
Functional check
Repair
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
Cleaning and disinfection
Fault
Cause
Remedy
Ball sticks after
flow has been
terminated
Static charge
Disconnect
flowmeter and
set down
briefly
Soiling inside
the flow tube
Replace flow
tube, call
DrägerService
Cracks and/or
clouding of the
pressure dome
Incorrect
disinfectant or
exposure to
sunlight
Replace the
pressure
dome. Call
DrägerService
Leakage
between
flowmeter and
connected
accessories
Missing or
worn sealing
ring at the gas
outlet
Call
DrägerService
WARNING
Risk of infection
Users and service personnel can become
infected with pathogenic germs.
Disinfect and clean device or device parts
before any maintenance measures and also
before returning the medical device for repair.
Concept
Definition
Maintenance
All measures (inspection,
preventive maintenance, repair)
intended to maintain and restore
the functional condition of a
medical device
Inspection
Measures intended to determine
and assess the actual state of a
medical device
Preventive
maintenance
Recurrent specified measures
intended to maintain the functional
condition of a medical device
Repair
Measures intended to restore the
functional condition of a medical
device after a device malfunction
Checks
Interval Personnel
responsible
Inspection
Every
5 years
Service personnel
WARNING
Risk of infection
Reusable products must be reprocessed. Oth-
erwise there is an increased risk of infection
and the functioning of the products may be im-
paired.
– Comply with the hospital’s hygiene regula-
tions.
– Use validated procedures for reprocessing.
– Reprocess reusable products before each
use.
– Observe the manufacturer’s instructions on
cleaning agents and disinfectants.
WARNING
Risk due to faulty products
Reusable products may show signs of wear,
such as cracks, deformations, discolorations,
or peeling.
Check products for signs of wear and replace if
necessary.
Содержание MP04551
Страница 74: ...74 75 76 76 76 77 77 77 78 78 78 78 79 80 80 80 81 82 82 ru...
Страница 75: ...75 20 REF SN 76 Dr ger Dr gerService Dr ger Dr ger 81 O2...
Страница 76: ...76 Dr ger AIR A DIN B C D E F BS G NF H JIS I J K NIST L O2 00437050 x bar H G F E A D C B I J E K L...
Страница 77: ...77 1 Drger D 2 K NIST 3 I 4 A F G H J E A SpO2 e 20 00337050 A...
Страница 79: ...79 1 2 3 Dr ger Dr gerService Dr ger 5...
Страница 147: ...147 p 148 149 149 149 150 150 150 150 151 151 151 152 152 152 152 153 154 154 bg...
Страница 148: ...148 20 mT REF SN 149 Dr ger Dr gerService Dr gerService Dr ger Dr ger 153 O2...
Страница 149: ...149 Dr ger Air A DIN B C D E F G H I J K NIST L 1 Dr ger D 2 NIST K O2 00437050 x bar H G F E A D C B I J E K L...
Страница 151: ...151 1 2 3 1 2 3 Drger Dr gerService Dr gerService Dr gerService Dr gerService 5...
Страница 155: ...155 156 157 157 157 158 158 158 158 159 159 159 160 160 160 160 161 162 162 el...
Страница 156: ...156 20 mT REF SN 157 Dr ger Dr gerService Dr gerService Dr ger Dr ger 161 O2...
Страница 159: ...159 1 2 3 1 2 CS 3 Dr ger Dr gerService Dr ger Dr gerService Dr gerService Dr gerService 5...
Страница 171: ...171 172 172 173 173 173 173 174 174 174 174 175 175 175 175 176 176 177 177 zh...
Страница 172: ...172 Dr ger 20 mT REF SN 172 Dr ger Dr gerService Dr gerService Dr ger Dr ger 176 O2 O2...
Страница 178: ...178 179 179 180 180 180 180 181 181 181 181 181 182 182 182 183 183 184 184 ko...
Страница 179: ...179 Dr ger 20 mT REF SN 179 Dr ger Dr gerService Dr gerService Dr ger 183 O2 O2...
Страница 185: ...185 186 187 187 187 188 188 188 188 188 189 189 189 190 190 190 191 191 192 ja...
Страница 186: ...186 Dr ger 20 mT REF SN 187 Dr ger Dr gerService Dr gerService Dr ger Dr ger 191 O2...
Страница 189: ...189 1 2 CS 3 Dr ger Dr gerService Dr ger Dr gerService Dr gerService Dr gerService 5 1...
Страница 193: ...This page intentionally left blank Instructions for use Flowmeter 193...