Instructions for use Movita
33
Troubleshooting
Movita lift: Emergency operation of the system
In the event of a power supply failure, the lift system
of the Movita lift version can also be moved without
using the electrically-operated height adjustment.
1
Take off the cover (A) (see assembly
instructions).
2
Actuate the motor shaft (B) using a 7 mm
socket wrench.
Insufficient gas supply, hissing
sound.
Pressurized gas is escaping from
inside the ceiling supply unit.
Notify hospital maintenance or
DrägerService to check the gas
supply. Connected devices must
also be taken into consideration.
Lift system: LED indicator is not lit
green:
Board/System without power,
cable connection interrupted.
Check power supply and fuses
(hospital maintenance).
LED indicator defective.
Replace control panel
(DrägerService).
Defective fuses on the board.
Replace fuses
(DrägerService).
Lift system does not respond
when actuated.
System without power, cable
connection interrupted.
Check power supply and fuses
(hospital maintenance).
Motor control defective.
Inform DrägerService.
Incorrect setting position of the
keyswitch.
Docked medical device and
position of the keyswitch do not
correspond.
Contact service personnel.
Lift system does not respond to
operation of the
down arrow
key.
Perseus safety sensor triggered. Remove obstacle and operate
up arrow
key.
Fault
Cause
Remedy
044
A
B
NOTE
The system moves slowly due to the gear ratio.