Instructions for use Movita
27
Operation
Movita lift: Electrically-operated height adjustment
The keys for height adjustment are either on the
control panel on the media head or media column
or optionally on the cable remote control.
Moving the media column/media head
upwards
1
Push the
up arrow
key (A) and release it in the
desired position.
Moving the media column/media head
downwards
2
Push the
down arrow
key (B) and release it in
the desired position.
At the top and bottom limit positions, the lifting
motor switches off automatically.
Status indicator
An LED indicator in the middle of the control panel
(A) or on the control handles (B) shows the status
of the system:
– LED indicator is lit continuously green: Ready
for operation.
– LED indicator flashes green: One or more keys
are not working. Inform DrägerService.
– LED indicator flashes alternately green and red:
Combination of several faults. Inform
DrägerService.
– LED indicator off: Not ready for operation.
Maximum actuation time of the
electrically-operated height adjustment
The electrically-operated height adjustment for the
ceiling supply unit may be actuated for a maximum
of 30 seconds. After that, the ceiling supply unit
must not be actuated for 5 minutes.
CAUTION
Danger of damage to objects
Ceiling supply units, especially lift systems, can
collide with other ceiling supply units, associated
surgical lights, or room walls as a result of brisk or
careless operation. This can cause devices
placed on the shelves (e.g. monitors) to be
clamped under the lift arm.
Because of this, damage to the arm system can
result, which in an extreme case can lead to parts
of the arm system breaking.
033
A
B
A B
033
NOTE
Observing these times prevents damage to the
supply unit.
A
B