Instructions for use Movita
17
Assembly and preparation
1
Connect the device mains plug (A) to the power
sockets of the media head/media column.
Connected power plugs must not be covered by
workspace components such as shelves or
drawers.
2
Connect the potential equalization cable (B) to
the associated potential equalization socket.
The potential equalization cable must meet the
requirements of the IEC 60601-1 standard.
The potential equalization connection must be
made in order that all metallic conducting
surfaces are at the same potential.
Connecting the probe for medical gas
supply
Park position
A
Push the probe into the gas terminal unit until it
engages for the first time (park position).
Operating position
B
Push the probe in until it engages for the
second time.
Interrupting the gas supply
A
Apply light pressure to the release bushing. The
probe will return to the park position.
Fully disconnecting the plug
B
Push the release bushing with more force and
at the same time remove the probe from the gas
terminal unit.
005
WARNING
Increased fire risk
Gas terminal units must not come into contact
with oil, grease or flammable liquids.
NOTE
The following instructions apply only to Dräger DIN
gas terminal units.